Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Biomet, Inc.: XP-XP Tibial Tray - Interlok 59 mm Item # 195750

Recall number
Z-2527-2019
Recalling firm
Zimmer Biomet, Inc.
Product
XP-XP Tibial Tray - Interlok 59 mm Item # 195750
Status
Terminated
Initiated
2019-08-15
Posted
not on file
Terminated
2021-04-16
Reason
The locking bar not fully engaging
Root cause
Manufacturing material removal
Action
Risk Manager Responsibilities: 1.Review this notification and ensure that affected personnel are aware of the contents. 2.If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3.Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form will be returned even if you do not have affected products at your facility. 4.Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5.If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Surgeon Responsibilities: 1.Review this notification for awareness of the contents. 2.There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 3.A visual and audible confirmation should be made to ensure complete locking bar insertion. 4.Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. 5.Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 6.If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-
Quantity
50 units
Distribution
State NY IN OH MI GA SC FL MO WI MN ND SD NE NJ PA MD IL IA CA VA NC DC WV AR AL MS AL MI WA OR ID LA FL KS TN MA KY OK TX TX CT AZ NV NM MT WY CO UT DC WY Country: KOREA CHINA CANADA AUSTRALIA JAPAN EMEA SWEDEN ITALY
Lot, serial or model codes
Lot Number 206540 057180 206590 206580 206550 030800 206560 457940 457880 457930 457970 457950
510(k) numbers
["K153657"]
PMA numbers
not on file
Product code
JWH
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register