Ab Sciex: AB SCIEX 3200MD QTRAP LC/MS/MS System Mass Spectrometer for In-Vitro Diagnostic Use. Instrument Part Number (REF): 5024500 3200MD series and 4500MD series are mass spectrometers intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an electrical and magnetic field according to their mass. It is intended for in vitro diagnostic purposes. For in vitro diagnostic use.
- Recall number
- Z-2527-2016
- Recalling firm
- Ab Sciex
- Product
- AB SCIEX 3200MD QTRAP LC/MS/MS System Mass Spectrometer for In-Vitro Diagnostic Use. Instrument Part Number (REF): 5024500 3200MD series and 4500MD series are mass spectrometers intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an electrical and magnetic field according to their mass. It is intended for in vitro diagnostic purposes. For in vitro diagnostic use.
- Status
- Terminated
- Initiated
- 2016-06-16
- Posted
- 2016-08-12
- Terminated
- 2017-03-02
- Reason
- MultiQuant MD software where under certain conditions a user can be presented with incorrect quantitative results when using the Sum Multiple Ions feature.
- Root cause
- Software design
- Action
- AB SCIEX issued recall notification by letter dated June 16, 2016, to all affected customers. The letter identified the problem, provided temporary measures until a permanent software is available. A response form was to be completed and returned. questions regarding this notice contact SCI EX at +1 289 982 2712.
- Quantity
- 2 units
- Distribution
- Worldwide Distribution - US (Nationwide) Internationally to Canada, Austria, China, Netherlands, Norway, India, Korea, Germany, France, Switzerland, Ireland, Italy, Singapore, Russia, UK, Malaysia,
- Lot, serial or model codes
- Software Version: MultiQuant MD 3.0, MultiQuant MD 3.0.1 and MultiQuant MD 3.0.2 software
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- DOP
- Firm FEI
- 3008935545
- Firm city
- Framingham
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28