Skip to the content

FDA Medical Device Recalls (openFDA)

Customed, Inc: Breast Pack- FJG, code 900-3118, contains: (2) GOWN LARGE STANDARD SMS AAMIIII (2) TOWEL ABSORBENT 15" X 20" (1) PENCIL CAUTERY PUSH BOTTOM (1) SYRINGE BULB 60CC (1) TUBE SUCTION CONNECT~" X 12' (1) COUNTER NEEDLE & BLADE 20C FOAM/MAG (1) BOWL UTIL TIY QUART 32oz (1) COVER MAYO STAND REINFORCED (2) SKIN MARKER WITH RULER (2) RULER (1) COVER TABLE REINFORCED 50" X 90" (2) CUP DENTURE 8oz TEAL WITH LID (6) TOWEL CLOTH HUCK BLUE (2) SYRINGE 20CC WITHOUT NEEDLE LUER LOCK (1) YANKAUER SUCTION TUBE WITHOUT VENT (1) DRAPE BREAST/CHEST SMS (10) SPONGE LAP PREWASH 18" X 18" XRD (1) SHEET% DRAPE REINFORCED 60" X 77" (4) DRAPE UTILITY WITH TAPE (1) BLADE SURG. #15 CARBON STEEL (2) PR. GLOVE SURG. CUSTOMGUARD CHLOROPRENE # 6.5 POWDER FREE (2) PR. GLOVE SURG. CUSTOMGUARD CHLOROPRENE # 7 POWDER FREE (1) PR. GLOVE SURG. CUSTOMGUARD CHLOROPRENE # 7.5 POWDER FREE (10) GAUZE SPONGE 4" X 4" 16PLY (1) PK. STRIP STERI CLOSURE W' X 4" (1) NEEDLE SPINAL ANESTH 22G X 3Y2 (1) NEEDLE HYPODERMIC 22G X 1Y2 (1) IV PRIMARY ADD ADMINISTRAT SET (1) STOPCOCK 3WAY WSWIVEL MALE (1) SYRINGE 60ML WITHOUT NEEDLE LUER LOCK Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.

Recall number
Z-2527-2014
Recalling firm
Customed, Inc
Product
Breast Pack- FJG, code 900-3118, contains: (2) GOWN LARGE STANDARD SMS AAMIIII (2) TOWEL ABSORBENT 15" X 20" (1) PENCIL CAUTERY PUSH BOTTOM (1) SYRINGE BULB 60CC (1) TUBE SUCTION CONNECT~" X 12' (1) COUNTER NEEDLE & BLADE 20C FOAM/MAG (1) BOWL UTIL TIY QUART 32oz (1) COVER MAYO STAND REINFORCED (2) SKIN MARKER WITH RULER (2) RULER (1) COVER TABLE REINFORCED 50" X 90" (2) CUP DENTURE 8oz TEAL WITH LID (6) TOWEL CLOTH HUCK BLUE (2) SYRINGE 20CC WITHOUT NEEDLE LUER LOCK (1) YANKAUER SUCTION TUBE WITHOUT VENT (1) DRAPE BREAST/CHEST SMS (10) SPONGE LAP PREWASH 18" X 18" XRD (1) SHEET% DRAPE REINFORCED 60" X 77" (4) DRAPE UTILITY WITH TAPE (1) BLADE SURG. #15 CARBON STEEL (2) PR. GLOVE SURG. CUSTOMGUARD CHLOROPRENE # 6.5 POWDER FREE (2) PR. GLOVE SURG. CUSTOMGUARD CHLOROPRENE # 7 POWDER FREE (1) PR. GLOVE SURG. CUSTOMGUARD CHLOROPRENE # 7.5 POWDER FREE (10) GAUZE SPONGE 4" X 4" 16PLY (1) PK. STRIP STERI CLOSURE W' X 4" (1) NEEDLE SPINAL ANESTH 22G X 3Y2 (1) NEEDLE HYPODERMIC 22G X 1Y2 (1) IV PRIMARY ADD ADMINISTRAT SET (1) STOPCOCK 3WAY WSWIVEL MALE (1) SYRINGE 60ML WITHOUT NEEDLE LUER LOCK Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
Status
Terminated
Initiated
2014-05-20
Posted
2014-08-29
Terminated
2017-08-14
Reason
Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
Root cause
Package design/selection
Action
Customed Inc. sent a Recall Notification letter dated June 3, 2014 via email and followed by letter to all affected customers. The letter identified the affected products, problem and actions to be taken. The letter informed customers if the affected products are in their inventory to: 1. Remove products from sale (quarantine them) and cease distribution. 2. Complete and return the enclosed acknowledgement form to Customed Inc .Recall Coordinator. Affected products should be returned to Customed Inc for credit. Customers with questions were instructed to contact Customed Inc. Recall Coordinator at 1-787-801-0100 Ext. 7540
Quantity
2 lots/ 36 units
Distribution
Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
Lot, serial or model codes
900-3118; 2 lots: 131210806 140111440
510(k) numbers
not on file
PMA numbers
not on file
Product code
LRO
Firm FEI
2648727
Firm city
Fajardo
Firm state
not on file
Firm country
PR

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register