FDA Medical Device Recalls (openFDA)
Hamilton Medical, Inc.: Hamilton-T1, REF: 161006, Intensive Care and Transport Ventilator
- Recall number
- Z-2526-2023
- Recalling firm
- Hamilton Medical, Inc.
- Product
- Hamilton-T1, REF: 161006, Intensive Care and Transport Ventilator
- Status
- Open, Classified
- Initiated
- 2023-07-26
- Posted
- 2023-09-20
- Terminated
- not on file
- Reason
- Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.
- Root cause
- Nonconforming Material/Component
- Action
- On July 26, 2023, Hamilton Medical issued Urgent Medical Device Correction via E-Mail. Hamilton Medical asked consignees to take the following actions 1. Check for affected devices in your facility. Get in contact with your local distribution partner and get the devices serviced with high priority. 2. Please sign the enclosed/attached sheet to confirm that, as an End- Customer, you have received and duly noted this Medical Device Correction.
- Quantity
- 76 units
- Distribution
- Distribution US nationwide, including Puerto Rico. OUS: Not provided
- Lot, serial or model codes
- UDI: 07630002801850 & 07630002813532 / Affected Serial Numbers: 15929 17600 17661 17677 17698 17699 17741 17850 17877 17913 17925 17955 17957 17969 17975 17976 17982 17985 17986 17991 17993 17995 18009 18166 18183 18189 18191 18199 18224 18244 18251 18255 18256 18260 18263 18272 18273 18276 18288 18289 18305 18308 18333 18340 18349 18360 18363 18377 18378 18383 18414 18415 18419 18441 18451 18458 18461 18502 18537 18544 18550 18778 18784 19237 19290 19291 19292 19293 19294 19298 19304 19305 19470 19483 19490 26117
- 510(k) numbers
- ["K181216"]
- PMA numbers
- not on file
- Product code
- CBK
- Firm FEI
- 2937708
- Firm city
- Reno
- Firm state
- NV
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28