FDA Medical Device Recalls (openFDA)

Ventec Life Systems, Inc.: VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001

Recall number
Z-2525-2026
Recalling firm
Ventec Life Systems, Inc.
Product
VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001
Status
Open, Classified
Initiated
2026-05-08
Posted
2026-06-23
Terminated
not on file
Reason
Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of primary power and absence/depletion of backup power sources, which may result in delayed caregiver awareness of a hazardous condition.
Root cause
Vendor change control
Action
On 5/8/2026, recall notices were mailed to customers who were asked to do the following: 1) Immediately discontinue use of affected devices. 2) Inspect current stock, quarantine/segregate any unused affected products, remove affected devices from clinical service and segregate to prevent further use. 3) Use an unaffected ventilator for patient support. 4) If you have distributed or transferred affected devices to other facilities, caregivers, or end users, you must promptly notify them of this recall and provide a copy of this notification. 5) Confirm receipt of this notification letter by returning the enclosed Customer Response Form via email to Ventec-Recall@reacthealth.com Upon receipt of the Customer Response Form, firm will arrange for repair of existing device(s). Return the affected product to React Health If you have any questions, please contact firm at 844-698-6276.
Quantity
1
Distribution
US Nationwide distribution in the state of UT.
Lot, serial or model codes
REF/UDI-DI/Serial Number/Lot: PRT-00853-000/00855573007839/155391P/807138, PRT-00490-001/00855573007792/155391/804613
510(k) numbers
["K162877"]
PMA numbers
not on file
Product code
CBK
Firm FEI
3013095415
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register