Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: ARTIS Pheno, Fluoroscopic X-Ray system, Model Number 10849000

Recall number
Z-2524-2018
Recalling firm
Siemens Medical Solutions USA, Inc
Product
ARTIS Pheno, Fluoroscopic X-Ray system, Model Number 10849000
Status
Terminated
Initiated
2018-05-30
Posted
not on file
Terminated
2020-02-14
Reason
During manual LAO/RAO rotation at maximum speed or a combination of LAO/RAO and Cran/Caud rotation at maximum speed, the possibility exists for the safety monitoring within the system control to become activated. As a result, the C-arm stops the rotation and can drop up to 40 mm (1.6 ).
Root cause
Device Design
Action
An Urgent Medical Device Correction notification was distributed to customers on 5/30/18. The notification provided the following information to customers: Siemens will correct the error by inspecting all affected ARTIS Pheno systems. As an immediate measure to prevent movement stops during a C-arm rotation followed by a C-arm drop, the maximum rotation speeds will be reduced. This correction limits the maximum manual LAO/RAO rotation from 25/sec to 20/sec and Cran/Caud rotation from 20/sec to 10/sec. All other system parameters as well as 3D acquisition and Perivision/Peristepping remain unaffected. You may continue to use the ARTIS Pheno system while awaiting implementation of the corrective action. Our service organization will contact you shortly to arrange a date to perform this corrective action with Update Instruction AX004/18/S. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Following the corrective action, the cause is eliminated and a recurrence of the fault is prevented. Additionally, Siemens is developing a solution to restore the system back to the original settings. This solution will be distributed to all affected customers upon completion.
Quantity
40 in the United States
Distribution
The devices were distributed to the following US states: AL, CA, CO, FL, IL, KY, MA, MI, MN, MS, ME, NJ, NY, OH, TN, TX, UT, and VA.
Lot, serial or model codes
164033,  164071,  164034,  164095,  164057,  164054,  164055,  164080,  164072,  164088,  164048,  164094,  164068,  164019,  164013,  164073,  164025,  164026,  164027,  164060,  164083,  164021,  164066,  164075,  164074,  164084,  164086,  164098,  164028,  164053,  164063,  164064,  164024,  164049,  164018,  164022,  164061,  164100,  164097,  164011
510(k) numbers
["K163286"]
PMA numbers
not on file
Product code
OWB
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register