Skip to the content

FDA Medical Device Recalls (openFDA)

Aomori Olympus Co., Ltd.: Reusable distal attachments Models: MH-462, MH-463, MH-464, MH-465, MH-466, MH-483, MH-587, MH-588(US), MH-589, MH-590, MH-591, MH-592, MH-593, MH-594, MH-595, MH-596, MH-597, MH-598, MAJ-289, MAJ-290, MAJ-291, MAJ-292, MAJ-293, MAJ-294, MAJ-295, MAJ-296, MAJ-297, MAJ-663, MAJ-664, MAJ-665, MAJ-666

Recall number
Z-2523-2023
Recalling firm
Aomori Olympus Co., Ltd.
Product
Reusable distal attachments Models: MH-462, MH-463, MH-464, MH-465, MH-466, MH-483, MH-587, MH-588(US), MH-589, MH-590, MH-591, MH-592, MH-593, MH-594, MH-595, MH-596, MH-597, MH-598, MAJ-289, MAJ-290, MAJ-291, MAJ-292, MAJ-293, MAJ-294, MAJ-295, MAJ-296, MAJ-297, MAJ-663, MAJ-664, MAJ-665, MAJ-666
Status
Open, Classified
Initiated
2023-07-27
Posted
2023-09-07
Terminated
not on file
Reason
The Distal Attachment contain di-2-ethylhexyl phthalate (DEHP), which exceeds the permissible exposure level calculated according to ISO 10993-17: 2002
Root cause
Nonconforming Material/Component
Action
Olympus issued Urgent Medical Device Recall Letter dated July 27, 2023 to ENDOSCOPY DEPARTMENT, RISK MANAGEMENT. Letter states reason for recall, health risk and action to take: Immediately assess any product you have to identify the product number with affected lot listed in this communication, cease use and quarantine any affected products. The image below depicts the area where the lot number is identified. The lot# is on the Carton box and Pack. Please refer to Attachment 2 for the identification of the subject products. 2. Call your Olympus customer service representative at 1-800-848-9024, option 2. Olympus will issue a Return Material Authorization to return any affected product at no charge to you. Olympus will issue a credit to your facility for your affected product. 3. Olympus requests that you acknowledge receipt of this letter through the Olympus recall portal. a. Go to https://olympusamerica.com/recall b. Enter the file (recall) number: 0429 4. If you have further distributed this product, identify your customers, forward them this notification, and appropriately document your notification process. Olympus requests that you report complaints, including patient physical condition suspected to be affected by DEHP, to Olympus. Please report complaints to our Technical Assistance Center (TAC) at 1-800-848-9024 (option 1). If you require additional information, please contact Tara Safi, Field Corrective Actions Administrator at Tara.Safi@Olympus.com.
Quantity
1,252 units
Distribution
US Nationwide distribution.
Lot, serial or model codes
Model: UDI: MH-462 4953170047732 MH-463 4953170047756 MH-464 4953170047770 MH-465 4953170047794 MH-466 4953170047817 MH-483 N/A MH-587 4953170047855 MH-588 4953170364518 (US) MH-589 4953170047893 MH-590 N/A MH-591 4953170047930 MH-592 N/A MH-593 4953170047978 MH-594 4953170047992 MH-595 4953170048012 MH-596 N/A MH-597 4953170048050 MH-598 N/A MAJ-289 N/A MAJ-290 N/A MAJ-291 N/A MAJ-292 N/A MAJ-293 N/A MAJ-294 N/A MAJ-295 N/A MAJ-296 N/A MAJ-297 N/A MAJ-663 N/A MAJ-664 N/A MAJ-665 N/A MAJ-666 N/A DIstal Attachment- All lots
510(k) numbers
K984358
PMA numbers
not on file
Product code
FDS
Firm FEI
3003995201
Firm city
Kuroishi
Firm state
2 Chome
Firm country
Japan

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this source