GE Healthcare, LLC: Brivo XR385, model 5215463, Digital Diagnostic Radiographic System

Recall number
Z-2523-2018
Recalling firm
GE Healthcare, LLC
Product
Brivo XR385, model 5215463, Digital Diagnostic Radiographic System
Status
Terminated
Initiated
2016-09-16
Posted
2018-07-16
Terminated
2018-07-26
Reason
The placing of weight on the VRAD detector, of the Brivo XR385 digital radiographic system, can introduce image artifacts, which can result in patient exam retakes.
Root cause
Component change control
Action
An Important Electronic Product Radiation Warning letter was sent to consignees beginning 9/16/2016. The letter identified affected product, described the issue, and stated that GE will correct the devices without charge. Questions can be directed to 1-800-437-1171.
Quantity
58 (55 US, 3 OUS)
Distribution
AK, AL, AZ, CA, CT, FL, GA, HI, IA, IL, KS, KY, LA, MA, MD, MS, NJ, NM, OR, PA, TN, TX, VA, and Guam, and Canada.
Lot, serial or model codes
Mfg Lot or Serial # / System ID:  000A1E36N037X1 / 319668B385, 000A1E36N037X1 / GON4182034 , 000A1E38N026L1 / 504EJLKV1, 000A1E38N026N2 / 907983BRIVO, 000A1E38N026R1 / 316630BRIVO, 000A1E38N026U2 / 319754B385, 000A1E38N026V1 / 860678BRIVO, 000A1E44N025A1 / 201567RAD1, 000A1E44N025G1 / 409295BRIVOXR, 000A1E44N025K2 / 860658XR385, 000A1E44N025O1 / 214MHXR385, 000A1E44N025V1 / 402384NHIRAD, 000A1E44N025Y1 / 256ODCRAD1, 000A1E44N025Z1 / 760323BRIVO1, 000A1E4BN003V2 / 772335B385, 000A1E4CN012F2 / 270737RAD1, 000A1E4CN012G1 / 804281XR385, 000A1E4CN012O1 / 240707RAD, 000A1E4CN012Q1 / 305MMC385, 000A1E4CN012T1 / 615396DR, 000A1E4CN076H2 / 203422BRIVO, 000A1E4CN076M1 / 856291BRIVO, 000A1E4CN076N1 / 808AAOMBRIVO, 000A1E4CN076R2 / 850325BC385, 000A1E4CN077P1 / 780875RAD, 000A1E4CN077R2 / 905850XR1, 000A1E4CN077V2 / 210INWD385ER, 000A1E52N024C2 / 623561XR385, 000A1E57N059P2 / 240707RAD, 000a1E57N060K2 / 813873BR1, 000A1E57N060L2 / 661KMSXR2, 000A1E57N060T2 / 707464BRIVO, 000A1E57N060W1 / 325655RAD, 000A1E57N060X1 / 973292BRIVO, 000A1E57N060X2 / 815464XR385, 000A1E57N060Y1 / 413447BRIVO, 000A1E5AN058C1 / 813906DR1, 000A1E5AN058D2 / 805898BRIVO1, 000A1E5AN058J1 / 508862BRIVO, 000A1E5AN058J2 / 575627BR385, 000A1E5AN058L1 / 671FHPBRIVO, 000A1E5AN058O2 / 508778XR385, 000A1E5AN058W1 / 541789BRIVO, 000A1E5AN058Y2 / 954PALMSXR, 000A1E5AN058Y2 / AM14GBU59, 000A1E5AN058Z1 / 415526BRIVO, 000A1E5AN059K1 / 306625XR385, 000A1E5AN059M2 / 662647BRIVO, 000A1E5AN059V2 / 575542BR385, 000A1E5AN059Y2 / PROVXR385, 000A1E5CN061G1 / 505609B385, 000A1E5CN061J1 / 863419BRI, 000A1E5CN061L1 / 404616BRIVO, 000A1E5CN061M1 / 615220PBRIVO, 000A1E5CN061V1 / 856355XR385, 000A1E5CN062H1 / 217774SXR, A1E5AN058X1 / 508548XR385, A1E67N001A2 / SLP385.
510(k) numbers
["K103448"]
PMA numbers
not on file
Product code
KPR
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register