Skip to the content

FDA Medical Device Recalls (openFDA)

Multi-Med, Inc.: All Med Medical 20ga x 1" Winged Administration Set w/Y Site Model # M201Y Reorder # 50-2032

Recall number
Z-2519-2010
Recalling firm
Multi-Med, Inc.
Product
All Med Medical 20ga x 1" Winged Administration Set w/Y Site Model # M201Y Reorder # 50-2032
Status
Terminated
Initiated
2010-08-30
Posted
2010-09-29
Terminated
2013-01-31
Reason
Huber needles may core and result in port leakage or emboli being flushed our of the port.
Root cause
Device Design
Action
Multi-Med issued an Urgent: Medical Device Recall notification on 8/30/10 via fax and followed by hard copy letter. Users were advised of the problem and asked to immediately examine their inventory and quarantine the affected product. They were to also notify their customers of the recall. The firm stated in the letter that they would contact customers to determine the number of product that they had and the amount they expected from the field. The firm will coordinate a plan for the return of the affected product. Questions should be directed to Sue Starkey at 603-357-8733.
Quantity
31,261 units
Distribution
Worldwide Distribution -- USA and Mexico.
Lot, serial or model codes
Lot code: 5G055A 5J034A 5J036A 6H027A 6I010A 6J053A 6K027A 7J008A 8D025A 8F029A 8L043A 9A024A
510(k) numbers
["K891919"]
PMA numbers
not on file
Product code
FMI
Firm FEI
1221765
Firm city
Keene
Firm state
NH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register