Philips Ultrasound, Inc.: 3D9-3v Transducer, accessory to Philips Ultrasound Systems: EPIQ Elite; Affiniti 30, 50, and 70; ClearVue 850; HD15; iU22; and Compact 5000

Recall number
Z-2517-2023
Recalling firm
Philips Ultrasound, Inc.
Product
3D9-3v Transducer, accessory to Philips Ultrasound Systems: EPIQ Elite; Affiniti 30, 50, and 70; ClearVue 850; HD15; iU22; and Compact 5000
Status
Completed
Initiated
2023-06-30
Posted
2023-09-06
Terminated
not on file
Reason
Transducer, an ultrasound system accessory, consists of two parts that are bonded together that may come apart due to chassis bonding issue, which may pose risk to users and patients of tissue damage, and electric shock if the built-in safety measures to prevent electric were to also fail.
Root cause
Process control
Action
On 6/30/23, recall notices were sent to customers who were advised to do the following: 1) Stop usage and/or distribution of affected transducers and secure them in a location where they cannot inadvertently be used. 2) Circulate the notice to all users of this device so that they are aware of the issue. Retain this notice with your system(s) and ensure the letter is in a place likely to be seen/viewed. 3) Complete and return the response form via email to: ultrasound.corrections@philips.com Field Service Engineers will visit your site to provide a replacement device, and collect the affected product. Contact Customer Care Solutions Center if you need further information or support concerning this issue at 1-800-722-9377. On 7/24/23 additional recall notices were distributed to customers informing them that additional devices were found to be affected.
Quantity
449
Distribution
Worldwide - US Nationwide distribution including in the states of FL, MN, AZ, NC, CA, MO, OH, WV, AL, NH, IN, IL, VT, PA, NY, NE, TX, WI, MI, KY, NV, MA, SD, LA, CO, TN, AK, GA, WA, UT, NM, ME, AR, SC and the countries of Italy, China, Netherlands, New Zealand, Hungary, Australia, France, Thailand, C¿te D'Ivoire, Germany, Canada, Mexico, Spain, Saudi Arabia, Uzbekistan, Greece, Yemen.
Lot, serial or model codes
UDI-DI:00884838061545, Serial Numbers: F07VBZ, F083JP, F086Z5, F088QC, F08JPF, F08K4H, F08VLP, F08VLY, F08W1C, F08W1D, F08W1J, F08W1K, F08W1L, F08W1M, F08W1Q, F08WXP, F08WXY, F08WZY, F08X01, F08X02, F08X03, F08X04, F08Y2T, F08ZR3, F08ZR4, F08ZR5, F092BQ, F092BR, F092BV, F097QQ, F097QV, F097QW, F09BTB, F09BTC, F09BTD, F09BTG, F09BTH, F09BTN, F09BTP, F09C1G, F09C1H, F09C1L, F09C1M, F09F3R, F09F3T, F09F3V, F09F3X, F09F3Y, F09F3Z, F09F40, F09F41, F09F42, F09F43, F09F45, F09F48, F09F49, F09F4C, F09F4D, F09F4F, F09F4G, F09F4H, F09JT0, F09JT2, F09JT3, F09MH4, F09MH5, F09MH6, F09MH7, F09MH8, F09MH9, F09MHB, F09MHC, F09MHG, F09MHH, F09MHK, F09MHL, F09MHM, F09MHN, F09MHP, F09N4J, F09N4K, F09N4M, F09N4N, F09N4P, F09NLM, F09NLN, F09NLP, F09NLT, F09V25, F09V26, F09V27, F09V28, F09V29, F09V2C, F09V2D, F09V2F, F09V2G, F09V2H, F09V2J, F09VH6, F09VH7, F09VH9, F09VHB, F09VWL, F09VWM, F09VWN, F09VWP, F09VWQ, F09XY4, F09XY5, F09XY6, F09XY7, F09XY8, F09XYB, F09XYD, F09XYF, F09XYG, F09XYH, F09YQ7, F09YQ8, F09YQ9, F09YQB, F0B16J, F0B16K, F0B16M, F0B16P, F0B1K5, F0B1K6, F0B1K7, F0B1K8, F0B1K9, F0B5KP, F0B5KQ, F0B5KR, F0B5KT, F0B5KV, F0B5KW, F0B6K9, F0B6KB, F0B6KC, F0B6KD, F0B6KF, F0B6KG, F0B7ND, F0B7NF, F0B7NG, F0B7NH, F0B7NJ, F0B7NK, F0B7NM, F0B7NN, F0B7NQ, F0B7NT, F0B7NX, F0B7NY, F0B7P0, F0B7P4, F0B7P5, F0B7X9, F0B7XC, F0B7XD, F0B7XG, F0B8G7, F0B8G8, F0B8G9, F0B8GB, F0B95D, F0B95G, F0BB01, F0BB02, F0BB03, F0BBGT, F0BBGX, F0BBGY, F0BBGZ, F0BCFQ, F0BCFR, F0BDMQ, F0BDMR, F0BDMT, F0BDMV, F0BFK8, F0BFKB, F0BFKC, F0BFXH, F0BFXJ, F0BFXK, F0BFXM, F0BFXN, F0BFXQ, F0BFXV, F0BHWP, F0BHWQ, F0BHWR, F0BHWT, F0BHWV, F0BHWW, F0BHZP, F0BHZR, F0BHZT, F0BHZV, F0BHZW, F0BHZZ, F0BJ00, F0BJ01, F0BJ9X, F0BJ9Y, F0BJB0, F0BJB1, F0BJB2, F0BJYQ, F0BJYR, F0BL03, F0BL06, F0BMC1, F0BMC4, F0BMC5, F0BMC6, F0BN4N, F0BN4T, F0BN4V, F0BR78, F0BR79, F0BR7B, F0BR7D, F0BRMH, F0BRMJ, F0BT87, F0BT88, F0BT8B, F0BT8C, F0BVPB, F0BVPC, F0BVPF, F0BW00, F0BW01, F0BW02, F0BW03, F0BW04, F0BW05, F0BXCT, F0BXCV, F0BXCW, F0BXCX, F0BXCY, F0BXT4, F0BXT5, F0BXT7, F0C3W1, F0C4DJ, B3TBRW, F01W12, F01ZG6, F01ZZ5, F024G2, F03Q35, F04DX8, F04DXC, F04DXD, F04F28, F04F29, F04F2G, F04F2H, F04F2L, F07GY0, F07XRT, F07XRV, F07XRX, F07Y4Q, F09NLQ, F09NLR, F0B95H, F0BBGW, B3PZBK, F03561, F03JJ7, F054CP, F08Y2Z, F05VJ4, F07QBQ, F021MP, F02WLR, F01GWF, B2ZHFB, B3PYF8, F00MC6, F01QCD, F03Z3Q, F05Q7Q, F09BTL, B2ZHX2, F04655, B3HX01, F09N4L, F026YN, F05Y8M, F045L3, B2ZVMB, F09BTM, F0BCFM, F0BB04, F0BFK9, F097QP, F061X4, F09F4B, F03K58, F08ZR1, F011M1, F01MPH, F02MX4, F01QCL, F06TCC, B35BYY, F09F3W, F0295R, F05DHN, F00XH9, F09XYC, B3Q08H, B1MKLW, B3HMKL, F009V6, F045L6, F06WH0, B3L4N2, F04F24, F01MV2, F0BT89, F0BVP9, B3FT4V, F04F26, F09F47, B3FH7X, F05J7K, F09JT4, B0WH4P, B3HY5P, B35JQ4, B3HLN6, F05PR5, F09MHD, F00Q08, F027TG, F049Y9, F09YQ6, F00MCQ, F00Z9K, F04Y32, F09F44, F0152T, F03LBN, B32MKY, B2JFNX, B3Q5WZ, B3L5BB, F0568H, B3FTPN, F00NQB, B2RVM9, B3LFRX, F009V5, B3LFT2, B3TCCV, F03MCP, B3L4FD, B1YQC1, F081WX, F05PR8, B3L4LZ, F033JD, F0B16N, B3LFJF, F02Y1L, F0BR7F, F0723Y, B2867Q, B3HX2C, F0152P, F07D6G, B3L4QJ, B2ZHB2, F009V0, B18NV8, F04444, B23M55, B3HWZT, B2ZR05, B32MK3, F02RND, F09C1J, F0B7NZ, F0B7P1, F0B7P3, F0BJ02, F0BL05, F0BL07, B25331, B3LDH7, F0603X, F097QR, F097QT, F09BTF, F09BTJ, F09BTK, F09C1K, F09F46, F09JT1, F09VH8, F0B7NL, F0B7NP, F0B7NR, F0B7NW, F0B7P2, F0B9ZZ, F0BB00, F0BBGV, F0BCFL, F0BCFN, F0BCFP, F0BDMW, F0BDMX, F0BFKD, F0BFXL, F0BFXP, F0BFXT, F0BFXW, F0BHZQ, F0BHZY, F0BJ9Z, F0BJYV, F0BMC3, B0T4K4, B2JDYH, B2PWBT, B2RL1T, B35JLT, B3TCYD, F01X0Z, F04TQ8, F05QVR, F06WH6, F07D6C, F09V2B, F0B7XB, F0B7XF, F0B8G6, F0B95B, F0B95C, F0BFK7
510(k) numbers
["K042540","K043535","K081661","K120321","K132304","K222648"]
PMA numbers
not on file
Product code
ITX
Firm FEI
3019216
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register