Skip to the content

FDA Medical Device Recalls (openFDA)

PREMIA SPINE LTD: TOPS Inserter, part of the TOPS System Instrument Set used for implantation of the TOPS System. Model Number: 82889.

Recall number
Z-2516-2025
Recalling firm
PREMIA SPINE LTD
Product
TOPS Inserter, part of the TOPS System Instrument Set used for implantation of the TOPS System. Model Number: 82889.
Status
Open, Classified
Initiated
2025-07-16
Posted
2025-09-04
Terminated
not on file
Reason
Potential for missing pins at tip of inserter.
Root cause
Process change control
Action
On 07/16/2025, Premia Spine Ltd. notified Premia Spine Inc. via phone call. It was requested to quarantine any suspected reworked instruments that are at the warehouse and to initiate a field action and return the affected units of the TOPS Inserter that underwent rework. No customers (hospitals/physicians) were asked to quarantine or return units as all affected units were within Premia Spine's control, however one surgeon was notified, on about 07/31/2025, of the issue and was requested to review the surgical procedure, analyze relevant patient data, including x-ray, and perform additional actions if needed to confirm the pin has not remained in the patient's body.
Quantity
30 units
Distribution
US Nationwide distribution in the state of Connecticut.
Lot, serial or model codes
Model Number: 82889. UDI-DI: 07290115932638. Lot Numbers: RS20220546-1, RS20222162-1, RS20171777-1, RS20171884-1, RS0431-13B-1, 20152764-1, RS0431-02-1.
510(k) numbers
not on file
PMA numbers
["P220002"]
Product code
QWK
Firm FEI
3012401682
Firm city
Netanya
Firm state
Ramat Poleg
Firm country
Israel

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register