FDA Medical Device Recalls (openFDA)
Bio-Rad Laboratories, Inc.: Platelia Toxo IgM Catalog Number 26211 UDI Code: 03610520005552 The Platelia Toxo IgM assay is a semi-quantitative immunocapture enzyme immunoassay kit for the detection of IgM antibodies to Toxoplasma gondii in human serum, plasma or cord blood specimens.
- Recall number
- Z-2515-2025
- Recalling firm
- Bio-Rad Laboratories, Inc.
- Product
- Platelia Toxo IgM Catalog Number 26211 UDI Code: 03610520005552 The Platelia Toxo IgM assay is a semi-quantitative immunocapture enzyme immunoassay kit for the detection of IgM antibodies to Toxoplasma gondii in human serum, plasma or cord blood specimens.
- Status
- Open, Classified
- Initiated
- 2025-05-30
- Posted
- 2025-09-04
- Terminated
- not on file
- Reason
- Due to a risk of false positive results that could lead to unnecessary medical treatment.
- Root cause
- Process control
- Action
- On 05/30/2025, the firm sent via email an "URGENT MEDICAL DEVICE NOTIFICATION" to customers informing them that, as a result of an internal investigation, it was revealed that samples taken from a frozen negative control panel tested unexpectedly interpreted as equivocal or positive results. Customer are instructed to: . As the issue can alter product performance of the affected lots by reducing specificity and increasing the number of equivocal results, interpret positive results with caution and contact Bio-Rad Technical Support if you observe an unusual rate of false positive results. . Refer to the test limitations, as outlined in Section 9 Limitations of the Procedure, of the IFU 0002043 version 2023/11: Diagnosis of recent infection by T. gondii can only be established on the basis of a combination of clinical and serological data. The result of a single serum sample does not constitute sufficient proof for diagnosis of recent infection . . As noted in Section 8 Interpretation of Results, positive IgM results should be followed by confirmatory steps in accordance with CDC guidelines before initiating or continuing treatment. For questions or assistance, contact Bio-Rad Technical Support at 1-800-224-6723 or by email at TechSupportUSSD-Redmond@Bio-Rad.com
- Quantity
- 35 kits
- Distribution
- U.S. Nationwide distribution in the states of CA, CO, FL, NC and NV.
- Lot, serial or model codes
- Catalog Number: 26211 UDI-DI code: 03610520005552 Batch/Lot Number: 4L0054
- 510(k) numbers
- ["K090847"]
- PMA numbers
- not on file
- Product code
- LGD
- Firm FEI
- 3002701146
- Firm city
- Woodinville
- Firm state
- WA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28