Bio-Rad Laboratories, Inc.: Platelia Toxo IgM Catalog Number 26211 UDI Code: 03610520005552 The Platelia Toxo IgM assay is a semi-quantitative immunocapture enzyme immunoassay kit for the detection of IgM antibodies to Toxoplasma gondii in human serum, plasma or cord blood specimens.

Recall number
Z-2515-2025
Recalling firm
Bio-Rad Laboratories, Inc.
Product
Platelia Toxo IgM Catalog Number 26211 UDI Code: 03610520005552 The Platelia Toxo IgM assay is a semi-quantitative immunocapture enzyme immunoassay kit for the detection of IgM antibodies to Toxoplasma gondii in human serum, plasma or cord blood specimens.
Status
Open, Classified
Initiated
2025-05-30
Posted
2025-09-04
Terminated
not on file
Reason
Due to a risk of false positive results that could lead to unnecessary medical treatment.
Root cause
Process control
Action
On 05/30/2025, the firm sent via email an "URGENT MEDICAL DEVICE NOTIFICATION" to customers informing them that, as a result of an internal investigation, it was revealed that samples taken from a frozen negative control panel tested unexpectedly interpreted as equivocal or positive results. Customer are instructed to: . As the issue can alter product performance of the affected lots by reducing specificity and increasing the number of equivocal results, interpret positive results with caution and contact Bio-Rad Technical Support if you observe an unusual rate of false positive results. . Refer to the test limitations, as outlined in Section 9 Limitations of the Procedure, of the IFU 0002043 version 2023/11: Diagnosis of recent infection by T. gondii can only be established on the basis of a combination of clinical and serological data. The result of a single serum sample does not constitute sufficient proof for diagnosis of recent infection . . As noted in Section 8 Interpretation of Results, positive IgM results should be followed by confirmatory steps in accordance with CDC guidelines before initiating or continuing treatment. For questions or assistance, contact Bio-Rad Technical Support at 1-800-224-6723 or by email at TechSupportUSSD-Redmond@Bio-Rad.com
Quantity
35 kits
Distribution
U.S. Nationwide distribution in the states of CA, CO, FL, NC and NV.
Lot, serial or model codes
Catalog Number: 26211 UDI-DI code: 03610520005552 Batch/Lot Number: 4L0054
510(k) numbers
["K090847"]
PMA numbers
not on file
Product code
LGD
Firm FEI
3002701146
Firm city
Woodinville
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register