Skip to the content

FDA Medical Device Recalls (openFDA)

Aesculap Implant Systems LLC: Univation X System knee implant devices as follows: Catalog/Product: NO180Z AS UNIVATION XF FEMUR CEMENTED F1 RM; NO181Z AS UNIVATION XF FEMUR CEMENTED F2 RM; NO182Z AS UNIVATION XF FEMUR CEMENTED F3 RM; NO183Z AS UNIVATION XF FEMUR CEMENTED F4 RM; NO184Z AS UNIVATION XF FEMUR CEMENTED F5 RM; NO185Z AS UNIVATION XF FEMUR CEMENTED F1 LM; NO186Z AS UNIVATION XF FEMUR CEMENTED F2 LM; NO187Z AS UNIVATION XF FEMUR CEMENTED F3 LM; NO188Z AS UNIVATION XF FEMUR CEMENTED F4 LM; NO189Z AS UNIVATION XF FEMUR CEMENTED F5 LM

Recall number
Z-2515-2021
Recalling firm
Aesculap Implant Systems LLC
Product
Univation X System knee implant devices as follows: Catalog/Product: NO180Z AS UNIVATION XF FEMUR CEMENTED F1 RM; NO181Z AS UNIVATION XF FEMUR CEMENTED F2 RM; NO182Z AS UNIVATION XF FEMUR CEMENTED F3 RM; NO183Z AS UNIVATION XF FEMUR CEMENTED F4 RM; NO184Z AS UNIVATION XF FEMUR CEMENTED F5 RM; NO185Z AS UNIVATION XF FEMUR CEMENTED F1 LM; NO186Z AS UNIVATION XF FEMUR CEMENTED F2 LM; NO187Z AS UNIVATION XF FEMUR CEMENTED F3 LM; NO188Z AS UNIVATION XF FEMUR CEMENTED F4 LM; NO189Z AS UNIVATION XF FEMUR CEMENTED F5 LM
Status
Terminated
Initiated
2021-03-19
Posted
not on file
Terminated
2025-03-28
Reason
Malfunction- loosening of the implant resulting in a potential revision surgery
Root cause
Unknown/Undetermined by firm
Action
Aesulap issued An Urgent Medical Device Recall via USPS overnight to consignees on March 19, 2021. Letter states reason for recall, health risk and action to take: Immediately examine your inventory, identify and quarantine product related to this notification. In addition, if you have further distributed this product, please identify your customers and notify them at once of this Recall Notification. Your notification to your customers may be enhanced by including a copy of this notification letter. A representative of Aesculap will reach out to you to assist in the return of affected product. 1. Review the univation X Knee System Recall Notification in its entirety and ensure that all users in your organization and other concerned persons are informed about this notification. 2. Active customers with inventory at their site were previously instructed (December 29, 2020 univation X Knee System Field Safety Notification via certified mail) to hold and do not use this product. We are requesting at this time that the product be returned. 3. Action is required of you to sign and return the provided Acknowledgement Response Form identifying (if applicable) the amount of product to the email address listed below before April 16, 2021. If the acknowledgement form has not been received at this time, another Recall Notification Letter will be sent. (Attachment 1 Medical Device Recall Notification Form). Send To: qa-recalls@aesculapusa.com 4. Identify your current inventory for the affected product. a. If you have identified inventory in your facility, DO NOT DESTROY any affected product. Contact Customer Service (info noted below) for an RGR number and return label to return your affected product. AIS Customer Service Number 1-866-229-3002 SetWorld_orders.BBMUS_Service@aesculapimplants.com b. Once the RGR and return paperwork is provided, return a copy along with your affected product to the address listed below. Aesculap Implant Systems SET WORLD HFPU71821 QC Recalls
Quantity
not on file
Distribution
US Nationwide distribution.
Lot, serial or model codes
All lots
510(k) numbers
["K081293","K131167"]
PMA numbers
not on file
Product code
HRY
Firm FEI
2916714
Firm city
Center Valley
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register