FDA Medical Device Recalls (openFDA)
BD SWITZERLAND SARL: BD Connecta" BD Luer-Lok" 360 REF 394910 UDI-DI code: 00382903949106 BD Connecta" Stopcocks are used to administer fluids or medication through one or two different ports via an IV cannula or extension tube. The ports can be used to sample blood or for hemodynamic monitoring. When used with IV lipid nutritional products, the stopcock device can be used for up to 24 hours.
- Recall number
- Z-2514-2026
- Recalling firm
- BD SWITZERLAND SARL
- Product
- BD Connecta" BD Luer-Lok" 360 REF 394910 UDI-DI code: 00382903949106 BD Connecta" Stopcocks are used to administer fluids or medication through one or two different ports via an IV cannula or extension tube. The ports can be used to sample blood or for hemodynamic monitoring. When used with IV lipid nutritional products, the stopcock device can be used for up to 24 hours.
- Status
- Open, Classified
- Initiated
- 2026-05-21
- Posted
- 2026-06-18
- Terminated
- not on file
- Reason
- Due to stopcock issues when using the plastic luer-lok syringe, the connection at the Port C may over thread or continue to spin rather than tightening securely. This behavior has led some clinical users to perceive the connection as unstable.
- Root cause
- Device Design
- Action
- On 05/21/2026, the firm sent via email and FedEx an "URGENT: Medical Device Correction" Letter informing customers that, BD has identified an observed issue in how Port C of the BD Connecta Stopcocks operates during use. Reports received indicate that when using polypropylene plastic luer lock syringes, the connection at Port C may over thread the syringe and continue to spin. This behavior has led some clinical users to perceive the connection as unstable. Actions for Clinical Users: -Per current Instructions for Use: "Do not overtighten connections. Note: Resistance between the male luer connector and the BD Connecta Stopcock when rotating indicates the connection is fully engaged and secure. Instruction to follow: 1.Circulate this notice to all those who need to be aware within your organization or to any organization where the potentially affected products have been transferred. 2.Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification. 3.Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA-1088 (1-800-332-1088) Mail: MedWatch, HF-2, FDA, 5600 Fisher s Lane, Rockville, MD 20852-9787 4.Report any complaints experienced with the use of this product to BD via the North American Regional Complaint Center: Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaints when prompted Mon Fri 8:00am and 5:00pm CT Email: productcomplaints@bd.com For Questions: Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaints when prompted Mon Fri 8:00am and 5:00pm CT or Email: productcomplaints@bd.com
- Quantity
- 2,465,500 units
- Distribution
- U.S.: AK, AL, AR, AZ, CA, CO, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, and WV. O.U.S.: Japan
- Lot, serial or model codes
- Catalog/Model Number: 394910 UDI-DI code: 00382903949106 Lot Numbers: 4277968 4277973 4282451 4282461 4289801 4290582 4292904 4295848 4299949 4303841 4305108 4306700 5129635 5129642 5129649 5129650 5129651 5139250 5139251 5139252 5150181 5153940 5164061 5164062 5178078 5181442 5183810 5183816 5183824 5183828 5218287 5218291 5218293 5218299 5218302 5241476 5247729 5247731 5251098 5275601 5275602
- 510(k) numbers
- K974083
- PMA numbers
- not on file
- Product code
- FMG
- Firm FEI
- 3012798826
- Firm city
- Eysins
- Firm state
- not on file
- Firm country
- Switzerland
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28