FDA Medical Device Recalls (openFDA)
GE Healthcare, LLC: Centricity PACS-IW with Universal Viewer - Product Usage: Is a device that displays medical images (including mammograms) and data from various imaging sources.
- Recall number
- Z-2502-2021
- Recalling firm
- GE Healthcare, LLC
- Product
- Centricity PACS-IW with Universal Viewer - Product Usage: Is a device that displays medical images (including mammograms) and data from various imaging sources.
- Status
- Open, Classified
- Initiated
- 2021-08-06
- Posted
- 2021-09-22
- Terminated
- not on file
- Reason
- Image acquisition failures and synchronization failure with the Centricity Enterprise Archive
- Root cause
- Software design
- Action
- This issue will be resolved by sending an Urgent Medical Device Correction letter to customers with affected software versions on August 06, 2021. This letter informs customers of the issues and provides safety instructions. GE Healthcare will correct the issue through a product change. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Quantity
- 392 (219 US; 173 OUS) all products
- Distribution
- AK, AZ, CA, CO, DE, FL, GA, IA, IN, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TX, UT, WA, WI, WV and WY. Austria, Brazil, Canada, France, Germany, Ghana, India, Ireland, Israel, Italy, Kuwait, Netherlands, Panama, Poland, Russia, Saudi Arabia, Slovenia, South Africa, Spain, Switzerland, Turkey, United Kingdom and Uruguay.
- Lot, serial or model codes
- 2068177-001 (Centricity PACS-IW with Universal Viewer software versions 5.0.x with PACS-IW foundation) System ID: 083055248481416, 1-121967267, 1-429329281, DGC19607785, HC3062UV01, HCIT1401012UW, HCIT4331279UV, PACS-IW-127, PACS-IW-219, PACS-IW-2770, PACS-IW-323, PACS-IW-421, RICHMONDU14767331, RICHMONDU15531177N1, X1347906 / X1347909, X1347906 / X1347909, YP315021, ZA2412PAC21 and ZA2533PAC1
- 510(k) numbers
- ["K123174"]
- PMA numbers
- not on file
- Product code
- LLZ
- Firm FEI
- 2126677
- Firm city
- Waukesha
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28