Skip to the content

FDA Medical Device Recalls (openFDA)

Argon Medical Devices, Inc: Scorpion Portal Vein Access Set - Product Usage: used to gain access to the hepatic vein and guide a sharp puncture tool (0.040 Stylet or 17ga Needle) through the parenchyma.

Recall number
Z-2501-2021
Recalling firm
Argon Medical Devices, Inc
Product
Scorpion Portal Vein Access Set - Product Usage: used to gain access to the hepatic vein and guide a sharp puncture tool (0.040 Stylet or 17ga Needle) through the parenchyma.
Status
Open, Classified
Initiated
2021-07-08
Posted
not on file
Terminated
not on file
Reason
As a result of design changes, sheaths have exhibited cracking/breaking at the tips.
Root cause
Device Design
Action
On 07/08/2021, the firm's sales representative/clinical Specialist contacted via telephone to inform the single customer that the Recalling Firm was pausing evaluations of the affected device and was requesting return of the device. The customer was also informed that the Recalling Firm was in the process of implementing a few changes to improve upon the product. The product was place on hold at the customer's facility and the Recalling Firm's sales representative was picked up and returned to the Recalling Firm on 07/21/2021.
Quantity
20 units
Distribution
U.S. Nationwide distribution in the state of WA. O.U.S.: N/A
Lot, serial or model codes
Item Numbers: TPS005 (Stylet); TPS006 (Needle); UDI Code: 00886333225552 Lot Number: 11369609
510(k) numbers
["K202141"]
PMA numbers
not on file
Product code
DYB
Firm FEI
1625425
Firm city
Athens
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register