Skip to the content

FDA Medical Device Recalls (openFDA)

Medline Industries, LP: MEDLINE Medical Procedure Kits labeled as: 1) CAROTID NEURO PACK, Medline Kit Number/SKU DYNJ44926G; 2) VP SHUNT CDS, Medline Kit Number/SKU CDS983800J; 3) VP SHUNT CDS, Medline Kit Number/SKU CDS983800K.

Recall number
Z-2500-2026
Recalling firm
Medline Industries, LP
Product
MEDLINE Medical Procedure Kits labeled as: 1) CAROTID NEURO PACK, Medline Kit Number/SKU DYNJ44926G; 2) VP SHUNT CDS, Medline Kit Number/SKU CDS983800J; 3) VP SHUNT CDS, Medline Kit Number/SKU CDS983800K.
Status
Open, Classified
Initiated
2026-05-04
Posted
2026-06-17
Terminated
not on file
Reason
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Root cause
Nonconforming Material/Component
Action
Medline Industries issued two URGENT MEDICAL DEVICE RECALL notice on 5/4/2026 to its consignees, one to its catheter and catheter tray consignees, and the other to the kit consignees via USPS and email. Both notices explained the issue and potential risk. The catheter and catheter tray consignees were directed to respond to the notice and destroy the affected products. The kit consignees were directed to respond to the notice and apply additional labeling (over-labels) to affix to the kits which instruct the user to remove and destroy the affected component. Both notices direct distributors or those who further distributing the products to notify their customers. For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Quantity
342 units
Distribution
Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
Lot, serial or model codes
Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 24EBH848; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 24DBK995; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 23JBP023; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 23HBR067; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 23FBM330; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 23EBK061; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 22JBG042; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 22GBH864; Medline Kit Number/SKU DYNJ44926G: UDI/DI 10193489766554 (each), 40193489766555 (case), Lot Number 22DBT571; Medline Kit Number/SKU CDS983800J: UDI/DI 10195327684662 (each), 40195327684663 (case), Lot Number 25DMJ438; Medline Kit Number/SKU CDS983800J: UDI/DI 10195327684662 (each), 40195327684663 (case), Lot Number 25CMH178; Medline Kit Number/SKU CDS983800J: UDI/DI 10195327684662 (each), 40195327684663 (case), Lot Number 24IME720; Medline Kit Number/SKU CDS983800J: UDI/DI 10195327684662 (each), 40195327684663 (case), Lot Number 24HMG912; Medline Kit Number/SKU CDS983800K: UDI/DI 10198459518973 (each), 40198459518974 (case), Lot Number 26CMC503; Medline Kit Number/SKU CDS983800K: UDI/DI 10198459518973 (each), 40198459518974 (case), Lot Number 26AMB047; Medline Kit Number/SKU CDS983800K: UDI/DI 10198459518973 (each), 40198459518974 (case), Lot Number 25JME109.
510(k) numbers
not on file
PMA numbers
not on file
Product code
OJG
Firm FEI
1417592
Firm city
Northfield
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register