Inverness Medical Professional Diagnostics: UCG Slide Test; Rapid latex agglutination inhibition slide test for the detection of hCG in urine. Used in the diagnosis of pregnancy.

Recall number
Z-2500-2010
Recalling firm
Inverness Medical Professional Diagnostics
Product
UCG Slide Test; Rapid latex agglutination inhibition slide test for the detection of hCG in urine. Used in the diagnosis of pregnancy.
Status
Terminated
Initiated
2010-02-12
Posted
2010-09-25
Terminated
2012-05-14
Reason
Specific lots no longer meet their stated limit of detection claims of 2000 mIU/mL for the UCG-Slide test and 500 mIU/mL for the UCG-Beta Slide Monoclonal II test.
Root cause
Other
Action
The recall was initiated on February 12, 2010. Letters were sent by Fed Ex on 2/12/2010.
Quantity
1052 kits
Distribution
Distributed Nationwide and to Australia
Lot, serial or model codes
Code number 15L3; lot numbers 0115299, Exp. 4.12.2010; 0632299, Exp. 7/30/2010; 0811039, Exp. 10/20/2009; Code number 15L5; lot numbers 0135029, Exp. 3.4.2010; 0221099, Exp. 4/12/2010; 0311179, Exp 5/27/2010; 0634299, Exp. 7/30/2010 and 0814039, Exp. 10/20/2010; Code number 15L8: lot numbers: 0136029, Exp. 3/4/2010; 0220099, Exp. 04/12/2010; 0627299, Exp. 07/30/2010; 0817039, Exp. 10/20/2010.
510(k) numbers
not on file
PMA numbers
not on file
Product code
LCX
Firm FEI
3003764480
Firm city
Princeton
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register