Skip to the content

FDA Medical Device Recalls (openFDA)

Exactech, Inc.: Threaded Offset Cup Impactor - Product Usage: To assist the surgeon in the implantation of Novation hip system components according to a conventional technique for total hip replacement. Specifically, these instruments are reusable, surgically invasive, and intended for transient use.

Recall number
Z-2499-2019
Recalling firm
Exactech, Inc.
Product
Threaded Offset Cup Impactor - Product Usage: To assist the surgeon in the implantation of Novation hip system components according to a conventional technique for total hip replacement. Specifically, these instruments are reusable, surgically invasive, and intended for transient use.
Status
Terminated
Initiated
2019-07-24
Posted
2019-09-10
Terminated
2023-05-31
Reason
The threaded mechanism of the Threaded Offset Cup Impactor may be gritty and difficult to turn potentially due to entrapped blast media.
Root cause
Nonconforming Material/Component
Action
Exactech notified customers on about 07/24/2019 via "URGENT MEDICAL DEVICE RECALL" letter. Instructions included to immediately cease distribution or use of the affected devices, provide the information to customers if further distributed, identify and quarantine any affected devices, and complete and return the provided Recall Acknowledgement Form and Recall Inventory Response Form to Exactech via email at recalls@exac.com. On 09/09/2019, Exactech expanded the recall to include an additional lot. Customers were notified of the expansion on about 09/09/2019.
Quantity
22 devices
Distribution
US distribution to AR, CA, FL, IL, IN, NY, SC, TX.
Lot, serial or model codes
Catalog Number 181-02-01, Lot Numbers 103071001, UDI Number 10885862520623.  Expansion on 09/09/2019 to include Lot Number 8182001
510(k) numbers
not on file
PMA numbers
not on file
Product code
HWA
Firm FEI
1038671
Firm city
Gainesville
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register