Skip to the content

FDA Medical Device Recalls (openFDA)

Ra Medical Systems, Inc.: DABRA Laser (RA-308 Excimer Laser). The DABRA Laser and DABRA Catheter is an excimer laser ultraviolet light source and delivery system, collectively the DABRA Laser System". The system is designed to deliver the energy to the surgical site to ablate unwanted material. The optical conduit is used in conjunction with existing endovascular accessories and tools to complete a surgical procedure. The software is programmed onto an EEPROM chip, which is a component of the DABRA Excimer Laser RA-308 system. Catheter for crossing total occlusions in vascular system.

Recall number
Z-2497-2020
Recalling firm
Ra Medical Systems, Inc.
Product
DABRA Laser (RA-308 Excimer Laser). The DABRA Laser and DABRA Catheter is an excimer laser ultraviolet light source and delivery system, collectively the DABRA Laser System". The system is designed to deliver the energy to the surgical site to ablate unwanted material. The optical conduit is used in conjunction with existing endovascular accessories and tools to complete a surgical procedure. The software is programmed onto an EEPROM chip, which is a component of the DABRA Excimer Laser RA-308 system. Catheter for crossing total occlusions in vascular system.
Status
Terminated
Initiated
2020-01-23
Posted
2020-07-04
Terminated
2021-11-01
Reason
A software issue was identified which could result in user or patient injury, or may adversely impact laser performance. It is possible that an unintended release of laser light (radiation) may occur, injuring the user or patient.
Root cause
Software design
Action
The Urgent Medical Device Correction notification was sent as certified mail USPS letter to consignees on 1/31/2020. The requested actions are as follows: 1. Do not use the excimer laser until a service technician has installed the new version of software to correct the issue. 2. Complete and return the acknowledgement form, whether you have any of the affected product or not, indicating: * Whether you have the excimer laser in question at your facility, *If you do have the excimer laser, the physical address where the product is located; and, *The person to be contacted to schedule a service call to correct the issue. Should you have any questions or need further assistance call (442) 325-5044.
Quantity
79 units
Distribution
US: AL, AZ, CA, FL, GA, ID, IL, IN, KS, KY, LA, MI, MD, NC, NV, NY, OK, OR, PA, TN,TX, WI. OUS: Italy and Japan.
Lot, serial or model codes
Serial Number(s): RA00027 through RA00176;all DABRA excimer laser RA-308 product is impacted.
510(k) numbers
["K170349"]
PMA numbers
not on file
Product code
PDU
Firm FEI
3004217728
Firm city
Carlsbad
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register