Philips Medical Systems (Cleveland) Inc: Ingenuity Digital PET/CT System, Model 882446 with the following UDI's: (01)00884838059559(21)900004 (01)00884838059559(21)900005 (01)00884838059559(21)900006 (01)00884838059559(21)900007 (01)00884838059559(21)900008 (01)00884838059559(21)900009 (01)00884838059559(21)900010 (01)00884838059559(21)900011 (01)00884838059559(21)900012 (01)00884838059559(21)900013 (01)00884838059559(21)900014 (01)00884838059559(21)900015 (01)00884838059559(21)900016 (01)00884838059559(21)900017 (01)00884838059559(21)900018 (01)00884838059559(21)900019 (01)00884838059559(21)900020 (01)00884838059559(21)900021 (01)00884838059559(21)900022 (01)00884838059559(21)900023 (01)00884838059559(21)900024 (01)00884838059559(21)900025 (01)00884838059559(21)900026 (01)00884838059559(21)900027 (01)00884838059559(21)900028 (01)00884838059559(21)900029 (01)00884838059559(21)900030 (01)00884838059559(21)900031 (01)00884838059559(21)900032 (01)00884838059559(21)900033 (01)00884838059559(21)900034 (01)00884838059559(21)900035 Product Usage: The system is intended to image the whole body, heart, brain, lung, gastrointestinal, bone, lymphatic, and other major organs for a wide range of patient types, sizes, and extent of diseases. Both subsystems can also be operated as fully functional, independent diagnostic tools, including application of the CT scanner for diagnosis and for use in radiation therapy planning.
- Recall number
- Z-2496-2018
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Product
- Ingenuity Digital PET/CT System, Model 882446 with the following UDI's: (01)00884838059559(21)900004 (01)00884838059559(21)900005 (01)00884838059559(21)900006 (01)00884838059559(21)900007 (01)00884838059559(21)900008 (01)00884838059559(21)900009 (01)00884838059559(21)900010 (01)00884838059559(21)900011 (01)00884838059559(21)900012 (01)00884838059559(21)900013 (01)00884838059559(21)900014 (01)00884838059559(21)900015 (01)00884838059559(21)900016 (01)00884838059559(21)900017 (01)00884838059559(21)900018 (01)00884838059559(21)900019 (01)00884838059559(21)900020 (01)00884838059559(21)900021 (01)00884838059559(21)900022 (01)00884838059559(21)900023 (01)00884838059559(21)900024 (01)00884838059559(21)900025 (01)00884838059559(21)900026 (01)00884838059559(21)900027 (01)00884838059559(21)900028 (01)00884838059559(21)900029 (01)00884838059559(21)900030 (01)00884838059559(21)900031 (01)00884838059559(21)900032 (01)00884838059559(21)900033 (01)00884838059559(21)900034 (01)00884838059559(21)900035 Product Usage: The system is intended to image the whole body, heart, brain, lung, gastrointestinal, bone, lymphatic, and other major organs for a wide range of patient types, sizes, and extent of diseases. Both subsystems can also be operated as fully functional, independent diagnostic tools, including application of the CT scanner for diagnosis and for use in radiation therapy planning.
- Status
- Terminated
- Initiated
- 2018-05-21
- Posted
- not on file
- Terminated
- 2020-04-10
- Reason
- The block used to connect the patient pallet to the couch may be assembled incorrectly. A screw may be misaligned in the patient pallet.
- Root cause
- Software design
- Action
- On May 21, 2018, Philips issued Customer Information notices to their clients. A Philips Field Service Engineer will contact customers directly to schedule the updates at your specific site. If you need any further information or support concerning this issue, please contact your local Philips representative. For SUPPORT in North America and Canada, contact the Customer Care Solutions Center 1-800-722-9377, follow prompts.
- Quantity
- 32 units
- Distribution
- Worldwide Distribution - US Nationwide in the states of CA, GA, LA, MD, MI, NY, OH, TN, TX, VT, Washington DC & WI Internationally - Canada, Mexico, Belgium, France, Israel, Luxembourg, Netherlands, Singapore, Spain & Switzerland
- Lot, serial or model codes
- Serial Numbers: 900004 900005 900006 900007 900008 900009 900010 900011 900012 900013 900014 900015 900016 900017 900018 900019 900020 900021 900022 900023 900024 900025 900026 900027 900028 900029 900030 900031 900032 900033 900034 900035
- 510(k) numbers
- ["K123599"]
- PMA numbers
- not on file
- Product code
- KPS
- Firm FEI
- 1525965
- Firm city
- Cleveland
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28