Radiometer Medical ApS: Blood gases (PCO2, PO2) and blood pH test system; ABL800 FLEX; Model Number 393-800; UDI (01)05700693938004. Model 393-800 is available in the following configurations: ABL835, ABL830, ABL827, ABL825, ABL825, ABL820, ABL817, ABL815, ABL810, ABL805 and ABL810 BG only and ABL800Basic. The FlexQ module is optional and can be added to all models and configurations.

Recall number
Z-2493-2020
Recalling firm
Radiometer Medical ApS
Product
Blood gases (PCO2, PO2) and blood pH test system; ABL800 FLEX; Model Number 393-800; UDI (01)05700693938004. Model 393-800 is available in the following configurations: ABL835, ABL830, ABL827, ABL825, ABL825, ABL820, ABL817, ABL815, ABL810, ABL805 and ABL810 BG only and ABL800Basic. The FlexQ module is optional and can be added to all models and configurations.
Status
Terminated
Initiated
2020-05-26
Posted
not on file
Terminated
2023-02-08
Reason
Potential risk of patient mix-up on analyzers due to software issues.
Root cause
Software design
Action
On 05/26/2020, the firm sent an Urgent: Medical Device Recall letter via email to all consignees; additionally the firm will send a hard copy letter FedEx 2nd day with tracking and proof of delivery.
Quantity
4,615 units
Distribution
US - AL, AR, AZ, BM, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV and WY. OUS - Australia, Austria, Canada, Czech Republic, China, Croatia, Denmark, France, Finland, Germany, Hungary, India, Italy, Japan, Netherlands, Norway, Poland, Russia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Turkey and United Kingdom
Lot, serial or model codes
Software versions 6.19 and below.
510(k) numbers
["K041874"]
PMA numbers
not on file
Product code
CHL
Firm FEI
3002807968
Firm city
Bronshoj
Firm state
not on file
Firm country
Denmark

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register