Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Medical Systems (Cleveland) Inc: Ingenuity TF PET/MR, Diagnostic Imaging System , Philips Medical Systems

Recall number
Z-2491-2015
Recalling firm
Philips Medical Systems (Cleveland) Inc
Product
Ingenuity TF PET/MR, Diagnostic Imaging System , Philips Medical Systems
Status
Terminated
Initiated
2015-07-23
Posted
2015-08-31
Terminated
2018-06-26
Reason
Philips Healthcare has become aware of a problem in which the Ingenuity TF PET/MR PET Reconstruction Server (PRS) database may lock up after an MR acquisition is completed and before a PET acquisition is begun, even though no actual database lockup issue has been reported in the field-installed Ingenuity TF PET/MR systems.
Root cause
Software design
Action
On July 23, 2015 the firm sent Field Safety Notices to their customers.
Quantity
11
Distribution
Product was shipped to the following states: NY, OH & TX. Product was also shipped to the following countries: Finland, Germany, Hong Kong, Japan, Netherlands, South Korea, Spain & Switzerland.
Lot, serial or model codes
M/N 882380; S/N: 11005, 11006, 11007, 11008, 11009, 11010, 11011, 11012, 11013, 11014 & 24061.
510(k) numbers
["K103483"]
PMA numbers
not on file
Product code
OUO
Firm FEI
1525965
Firm city
Cleveland
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register