Skip to the content

FDA Medical Device Recalls (openFDA)

RAYSEARCH LABORATORIES AB: RayStation with the following product descriptions: 1. RayStation 10B, Software Version: 10.1.0.613. 2. RayStation 10B Service Pack 1, Software Version: 10.1.1.54. Product Description: Radiation Therapy Treatment Planning System.

Recall number
Z-2490-2025
Recalling firm
RAYSEARCH LABORATORIES AB
Product
RayStation with the following product descriptions: 1. RayStation 10B, Software Version: 10.1.0.613. 2. RayStation 10B Service Pack 1, Software Version: 10.1.1.54. Product Description: Radiation Therapy Treatment Planning System.
Status
Open, Classified
Initiated
2025-08-08
Posted
2025-09-02
Terminated
not on file
Reason
DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
Root cause
Under Investigation by firm
Action
RaySearch notified consignees on about 08/08/2025 via email. Consignees were informed that as temporary solution, affected objects can be re-exported via RayGateway to generate new UIDs. They were instructed to inform planning staff and all users about this workaround, inspect your product and identify all installed units with the above software version number(s), and complete and return the response form. The issue will be resolved in the next version of RayStation, scheduled for market release in December 2025 (subject to market clearance in some markets). If customers wish to continue using versions of RayStation affected by this notice, all users must maintain awareness of this notice. Alternatively, customers can choose to upgrade to the new version once it becomes available for clinical use.
Quantity
3
Distribution
Worldwide - US Nationwide distribution in the states of AL, CA, FL, GA, IL, IA, KY, LA, MD, MN, MS, NY, NC, OH, OK, SC, WA, WI and the countries of Canada, China, Colombia, France, Germany, Hong Kong S.A.R., China, Italy, Japan, Netherlands, Poland, Spain, Sweden, Switzerland, ENG, Thailand, Turkey, UK, Ukraine.
Lot, serial or model codes
1. Software Version: 10.1.0.613, UDI: 0735000201031020201216. 2. Software Version: 10.1.1.54, UDI: 0735000201047120220128. Expiration Date: 2027-02-01
510(k) numbers
["K210645"]
PMA numbers
not on file
Product code
MUJ
Firm FEI
3007774465
Firm city
Stockholm
Firm state
not on file
Firm country
Sweden

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register