Skip to the content

FDA Medical Device Recalls (openFDA)

Paragon 28, Inc.: Phantom Nail System, Drill, 4.6 x 300mm, Cannulated, 3/16" Sq. Connection p28 P99-110-4630

Recall number
Z-2488-2021
Recalling firm
Paragon 28, Inc.
Product
Phantom Nail System, Drill, 4.6 x 300mm, Cannulated, 3/16" Sq. Connection p28 P99-110-4630
Status
Terminated
Initiated
2021-07-30
Posted
not on file
Terminated
2021-12-15
Reason
A thin wall condition between the inner cannulation and the flutes of the drill which potentially could cause a fracture through the wall of the drill to the cannulation beneath.
Root cause
Process change control
Action
On 08/12/2021, the Recalling Firm emailed to its distributors an "Medical Device Recall" Notification Letter inform them that it was identified that there is a thin wall condition between the inner cannulation and the flutes of the drill. At worst case, the defect is a fracture through the wall of the drill to the cannulation beneath. This feature is potentially nonconforming in some units within the lot identified below. The distributor was instructed to: 1) Immediately check its inventory and hospitals/health care facility within their region and to identify, discontinue use, quarantine and not distribute any further the affected products. 2) Complete the Medical Device Recall Response Form and return to mkinnee@paragon28.com. The form is to be completed and returned even if no physical product is in their possession. 3) Immediately return affected devices along with a copy of the Medical Device Recall Response Form to the Recalling Firm at Paragon28, Attn: Recalls, RN-008, 14445 Grasslands Dr., Englewood, CO 80112. For questions or further assistance - contact the Recalling Firm at: email - mkinnee@paragon28.com or complaints@paragon28.com Phone - 949-280-7688
Quantity
14 drill bits
Distribution
U.S. Nationwide distribution in the states of CA, FL, GA, ID, KS, KY, LA, MI, MO, NC, NY, TX, VA, WA and WV. O.U.S.: N/A
Lot, serial or model codes
Model Number: P99-110-4630 UDI Code: (01)00889795056849(10)HN120550 Lot Number: HN120550
510(k) numbers
not on file
PMA numbers
not on file
Product code
HTW
Firm FEI
3008650117
Firm city
Englewood
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register