PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Wireless Foot Switch distributed with Philips Allura Xper and Azurion Interventional Fluoroscopic X-Ray Systems

Recall number
Z-2485-2023
Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product
Wireless Foot Switch distributed with Philips Allura Xper and Azurion Interventional Fluoroscopic X-Ray Systems
Status
Open, Classified
Initiated
2023-06-30
Posted
2023-08-30
Terminated
not on file
Reason
Loss of availability of the wireless foot switch during procedures.
Root cause
Under Investigation by firm
Action
Customers were sent an URGENT Medical Device Correction letter dated 6/23/2023 via USPS Certified Mail. Customers are instructed to ensure that the wired footswitch is always connected to the system in case the wireless foot switch becomes unavailable. Customers are instructed to download a copy of the addendum to the IFU for the wireless foot switch and the Quick Reference Card at Http://philips.com/doc_library, and distribute these materials to those who use the system and retain a copy with the system's IFU. These documents are to be placed in an area where they are likely to be seen/viewed with system documentation. Customers are to return the attached response form to Philips no later than 30 days from receipt of the notification to IGT_Recalls@philips.com. Philips will contact customers to schedule a Field Service Engineer visit to confirm that the wired foot switch is connected to affected systems, perform a check of the charger, and provide customers with a hardcopy of the addendum. Customers with questions can contact Philips Technical Support Line at 1-800-722-9377.
Quantity
1865 units
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Bosnia and Herzegovina, Brazil, Chile, China, Croatia, Czech Republic, Denmark, Egypt, Fiji, Finland, France, Germany, Greece, Hong Kong, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Korea, Republic of, Kosovo, Latvia, Lebanon, Libya, Malaysia, Maldives, Malta, Martinique, Mexico, Morocco, Netherlands, New Zealand, Norway, Oman, Palestine, State of, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Tanzania, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, and Yemen.
Lot, serial or model codes
Model No.: 722001, 722003, 722005, 722006, 722008, 722010, 722011, 722012, 722013, 722015, 722023, 722025, 722026, 722027, 722028, 722029, 722033, 722035, 722038, 722039, 722058, 722059, 722064, 722065, 722066, 722067, 722068, 722078, 722079; No UDI-DI; Lot No. 459800772231, 459800772233, 459800772261, 459800772263, 459800415531,459800415532, 459800415533, 459800415534, 459800415535, 459800415571, 459800415572, 459800415573, 459800415574, 459800415575.
510(k) numbers
["K161563","K200917"]
PMA numbers
not on file
Product code
OWB
Firm FEI
3002807880
Firm city
Best
Firm state
not on file
Firm country
Netherlands

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register