FDA Medical Device Recalls (openFDA)
Olympus Corporation of the Americas: ViziShot 2FLEX. Single Use Aspiration Needle. Model Number: NA-U403SX-4019. The ViziShot 2 FLEX tissue biopsy device is made up of four main parts: a handle, a sheath, a needle, and a stylet. The sheath and needle are attached to the handle, and the stylet, which is removable, is located inside the needle tube.
- Recall number
- Z-2484-2025
- Recalling firm
- Olympus Corporation of the Americas
- Product
- ViziShot 2FLEX. Single Use Aspiration Needle. Model Number: NA-U403SX-4019. The ViziShot 2 FLEX tissue biopsy device is made up of four main parts: a handle, a sheath, a needle, and a stylet. The sheath and needle are attached to the handle, and the stylet, which is removable, is located inside the needle tube.
- Status
- Open, Classified
- Initiated
- 2025-08-06
- Posted
- 2025-09-05
- Terminated
- not on file
- Reason
- Potential for undetected, deformed a-traumatic tips.
- Root cause
- Process design
- Action
- Olympus notified consignees on about 08/06/2025 via letter. Consignees were instructed to examine inventory and quarantine any affected units, ensure all users of the device carefully read the content of this notification, including the reinforced text from the IFU and the product removal information, arrange for the return of affected units, and acknowledge receipt of the recall notification via the Olympus web portal. Consignees were also instructed to notify customers if the affected units were further distributed. ***Update 12/31/2025*** URGENT MEDICAL DEVICE RECALL letters were sent to additional customers on January 9, 2026 with identical instructions. ***Edit 3/23/26: January 9, 2026 letters users were instructed to return all affected devices.
- Quantity
- 104,508 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Germany, Canada, Chile, China, Hong Kong, India, Japan, Singapore.
- Lot, serial or model codes
- Model Number: NA-U403SX-4019. UDI-DI: 00821925043060. Lot numbers: All lots manufactured after 7/29/2022.
- 510(k) numbers
- ["K193517"]
- PMA numbers
- not on file
- Product code
- KTI
- Firm FEI
- 2429304
- Firm city
- Center Valley
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28