Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Sustainability Solutions: Reprocessed HARMONIC Ace +7, 5mm Diameter Shears with Advanced Hemostasis (HARH), Item Number HARH36 Product Usage: The Reprocessed HARMONIC ACE+7, 5 mm Diameter Shears with Advanced Hemostasis are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers and steel scalpels in general, plastic, pediatric, gynecologic, urologic, thoracic, exposure to orthopedic structures (such as spine and joint space), sealing and transection of lymphatic vessels, and other open and endoscopic procedures. The instruments allow for the coagulation of vessels up to and including 7 mm in diameter, using the Advanced Hemostasis hand controls button.

Recall number
Z-2484-2018
Recalling firm
Stryker Sustainability Solutions
Product
Reprocessed HARMONIC Ace +7, 5mm Diameter Shears with Advanced Hemostasis (HARH), Item Number HARH36 Product Usage: The Reprocessed HARMONIC ACE+7, 5 mm Diameter Shears with Advanced Hemostasis are indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instruments can be used as an adjunct to or substitute for electrosurgery, lasers and steel scalpels in general, plastic, pediatric, gynecologic, urologic, thoracic, exposure to orthopedic structures (such as spine and joint space), sealing and transection of lymphatic vessels, and other open and endoscopic procedures. The instruments allow for the coagulation of vessels up to and including 7 mm in diameter, using the Advanced Hemostasis hand controls button.
Status
Terminated
Initiated
2018-05-08
Posted
not on file
Terminated
2019-07-29
Reason
Stryker Sustainability Solutions has received an increase in reports indicating that the affected Reprocessed HARH36 devices may display an error code (No Instrument Uses Remaining) upon initial connection to the generator. When this error code is present, the device is not able to be used.
Root cause
Device Design
Action
Firm notified sales reps via email on 5/8/18 and instructed to have the customers complete response forms.
Quantity
180 units
Distribution
US Nationwide Distribution in the states of PA, FL, MD, MA, CA, NY, NC, TX, MT, AL, LA, TN
Lot, serial or model codes
Lot number 7656512
510(k) numbers
["K161693"]
PMA numbers
not on file
Product code
NLQ
Firm FEI
3009350690
Firm city
Tempe
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register