Abbott Laboratories, Inc: ARCHITECT i1000SR" Processing Module System For In Vitro Diagnostic Use, Abbott Laboratories Abbott Park, IL Product List Number, Part # or Model: 01L86-01 In Vitro Diagnostic use only. The Abbott ARCHITECT System is designed to perform automated: Chemistry tests, utilizing photometry and potentiometric technology, and Immunoassay tests, utilizing CMIA (chemiluminescent microparticle assay) detection technology.

Recall number
Z-2482-2010
Recalling firm
Abbott Laboratories, Inc
Product
ARCHITECT i1000SR" Processing Module System For In Vitro Diagnostic Use, Abbott Laboratories Abbott Park, IL Product List Number, Part # or Model: 01L86-01 In Vitro Diagnostic use only. The Abbott ARCHITECT System is designed to perform automated: Chemistry tests, utilizing photometry and potentiometric technology, and Immunoassay tests, utilizing CMIA (chemiluminescent microparticle assay) detection technology.
Status
Terminated
Initiated
2008-10-21
Posted
2010-09-23
Terminated
2012-04-13
Reason
During manufacturing, two components of the ARCHITECT i1000SR were reversed which could potentially cause falsely elevated patient results due to a sample-reagent and reagent-sample carryover.
Root cause
Mixed-up of materials/components
Action
ABBOTT issued Technical Service Bulletins to Field Service Representatives who then inspected every unit .
Quantity
404 units distributed worldwide (including US)
Distribution
Worldwide distribution:USA including states of: CO, DC, FL, GA, IL, LA, MA, MN, NY, OH, OK, TX, UT, and WV and countries including: Australia, Brazil, Bogota, Canada, China WFOTC, Columbia, Curacao (NETH ANT), Germany, Hong Kong, Japan, Mexico, Singapore, and Taiwan.
Lot, serial or model codes
Serial Numbers: (1st recall) i1SR01001-1224; 1226-1300; 1302-1306; 1308-1313; 1315-1316; 1318-1352; 1355-1363; 1365; 1376; 1381-1382  Extended to include serial numbers manufactured from 04/01/08 to 03/31/09: i1SR01021, 1026, 1032, 1040, 1036, 1043, 1059, 1065, 1069, 1071, 1071, 1074, 1082, 1094, 1106, 1107, 1109, 1111, 1112, 1114, 1117, 1119, 1121, 1123, 1128, 1133, 1135, 1138, 1139, 1158, 1162, 1167, 1171, 1173, 1177, 1183, 1186, 1188, 1189, 1198, 1200, 1201, 1205, 1209, 1211, 1217, 1225, 1230, 1243, 1246, 1249, 1250, 1260, 1262, 1272, 1277, 1279, 1284, 1286-1288, 1290, 1294, 1301, 1304, 1312, 1317, 1321, 1324-1327, 1332-1335, 1338, 1339, 1342-1345, 1347, 1348, 1351-1356, 1363-1368, 1370-1372, 1375, 1377-1387, 1389-1396, 1398-1410, 1412-1414, 1416-1441, 1443-1530, 1532-1536 (some of these #'s were removed whose field status was Demo/Training and instruments that had TSB 117-004 completed prior to Feb 19.
510(k) numbers
not on file
PMA numbers
not on file
Product code
JJE
Firm FEI
1628664
Firm city
Irving
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register