Skip to the content

FDA Medical Device Recalls (openFDA)

Terumo Cardiovascular Systems Corporation: Terumo Advanced Perfusion System 1, Catalog number 801764, 220/240V System 1 base, Terumo Cardiovascular System, Ann Arbor MI. (This product is not distributed in the US)

Recall number
Z-2481-2010
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Terumo Advanced Perfusion System 1, Catalog number 801764, 220/240V System 1 base, Terumo Cardiovascular System, Ann Arbor MI. (This product is not distributed in the US)
Status
Terminated
Initiated
2010-09-10
Posted
2010-09-23
Terminated
2011-12-20
Reason
The design document states that the oxygen sensor would operate for 300,000 hours but the supplier of the oxygen sensor indicated that it has a 100,000 hour warranty. The oxygen sensor is a consumable part that slowly depletes over time. At the end of its service life, a depleted oxygen sensor results in a gas system calibration failure or a message to the user prompting service for the gas syst
Root cause
Nonconforming Material/Component
Action
Terumo issued a URGENT MEDICAL DEVICE CORRECTION letter dated September 10, 2010, identifying the affected product, the issue prompting the correction, and what actions are to be taken by the customer. The firm included the addendum to the operators manual with the recall letter. The addendum recommended replacement of the oxygen sensor every six months instead of every 12 months. The addendum provided additional user instructions in the event of a calibration failure of the sensor. Customers are to complete and return the enclosed Customer Response Form to the firm. Terumo CVS Customer Service can be contacted by 1-800-521-2818 Monday through Friday, 8 AM 6 PM EST.
Quantity
1315
Distribution
Worldwide Distribution: USA, including the states of AL,AZ, CA, CO, CT, DC, FL, GA, HI, IA, IN, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WI, and WV, and the countries of Canada, Indonesia, China, Chile, UAE, Japan, Singapore, UK, Malaysia, Taiwan, Belgium, Korea, Phiippines and Thailand.
Lot, serial or model codes
Serial numbers 0001-1315.
510(k) numbers
["K022947"]
PMA numbers
not on file
Product code
DTQ
Firm FEI
1828100
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register