Skip to the content

FDA Medical Device Recalls (openFDA)

Draeger Medical Systems, Inc.: Fabius Tiro Anesthesia Machine, OR/Anesthesia, Draeger Medical, Inc., 3135 Quarry Road, Telford, PA 18969 Compact anesthesia system that offers the full spectrum of anesthesia ventilation to military anesthesiologists/ CRNAs in the field of operations. Can be used in a variety of mobile and stationary military applications where general anesthesia is required.

Recall number
Z-2479-2010
Recalling firm
Draeger Medical Systems, Inc.
Product
Fabius Tiro Anesthesia Machine, OR/Anesthesia, Draeger Medical, Inc., 3135 Quarry Road, Telford, PA 18969 Compact anesthesia system that offers the full spectrum of anesthesia ventilation to military anesthesiologists/ CRNAs in the field of operations. Can be used in a variety of mobile and stationary military applications where general anesthesia is required.
Status
Terminated
Initiated
2010-08-26
Posted
2010-09-23
Terminated
2011-08-23
Reason
potential failures to Electri-Cord Manufacturing AC power cord
Root cause
Other
Action
Draeger Medical, Inc., sent an Urgent - Medical Device Recall letter dated August 2010 to all customers. The letter described the product, the problem, and actions to be taken by the customer. Customers were instructed to: 1) Immediately take the cord out of service if the plug has bent or cracked prongs, burned plastic, or shows signs of excessive wear. 2) Monitor the affected cords regularly and be mindful of excessive wear and tear, misuse, or abuse until the affected cord is replaced. 3) Customers were also asked to complete and return the enclosed Customer Power Cord Form via mail: Draeger Medical, Inc., 3135 Quarry Road, Telford, PA 18969; fax: 1-215-721-5808; or e-mail: kathy.benedetto@draeger.com. 4) Disable the cords by cutting off the plug and dispose of the cord in accordance with their facility's disposal procedure when affected power cords are removed. NOTE: The firm will send replacement power cords free of charge. Draeger will only replace power cords originally provided by Draeger (identified with the white flag/part number 4117266. If you have any questions regarding this letter please call 1-800-543-5047 (press 1 at the first prompt, 2 at the second prompt and then 32349#).
Quantity
up to 1973 products (GS & Tiro)
Distribution
Worldwide Distribution: USA and countries of Brazil, Canada, Columbia, Costa Rica, Cuba, Ecuador, Japan, Lebanon, Nicaragua, Mexico, Panama, Philippines, Saudi Arabia, Taiwan, and Venezuela
Lot, serial or model codes
catalog #8606000
510(k) numbers
["K042419"]
PMA numbers
not on file
Product code
BSZ
Firm FEI
2510954
Firm city
Telford
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register