Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics Inc: epoc BGEM Crea Test Card with epoc Host SW v3.37.3, epoc NXS SW v4.10.6, Sensor Configuration 41.1

Recall number
Z-2477-2023
Recalling firm
Siemens Healthcare Diagnostics Inc
Product
epoc BGEM Crea Test Card with epoc Host SW v3.37.3, epoc NXS SW v4.10.6, Sensor Configuration 41.1
Status
Open, Classified
Initiated
2023-06-01
Posted
2023-08-28
Terminated
not on file
Reason
There is potential for discrepant high glucose results in samples with glucose results on the lower end of the reportable range.
Root cause
Under Investigation by firm
Action
Consignees were notified via an Urgent Medical Device Correction letter beginning June 5, 2023. Consignees were informed of the issue and were provided a workaround and actions to take. The next sensor configuration/software version, sensor configuration 42.1/epoc Host SW v3.38.2 /epoc NXS SW v4.11.11, is available to customers in Siemens Healthineers Document Library or the epoc Live Update Service (eLUS). This next sensor configuration, 42.1, mitigates the issue.
Quantity
473 ea
Distribution
US distribution to AL, AR, AZ, CA, CO, DC, DE, FL, IA, ID, IL, KS, KY, LA, MI, MN, MO, NC, ND, NH, NY, OH, OK, SC, SD, and TX OUS distribution to Canada, China, Colombia, Finland, India, Mexico, Norway, Sweden, United Arab Emirates, and Vietnam
Lot, serial or model codes
UDI-DI: 00809708072254 Siemens Material Number: 10736382 Lots: 03-23002-60 (expiry 19-Jun-2023) and 03-23056-50 (expiry 12-Aug-2023)
510(k) numbers
["K200107"]
PMA numbers
not on file
Product code
CHL
Firm FEI
3002637618
Firm city
Norwood
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register