FDA Medical Device Recalls (openFDA)
Roche Diagnostics Operations, Inc.: Cobas 8000 Modular Analyzer Series: cobas 8000 ISE module part Numbers : 05964075001 cobas c 701 module part Numbers : 05641489001 cobas c 702 module part Numbers : 06473245001
- Recall number
- Z-2477-2019
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Product
- Cobas 8000 Modular Analyzer Series: cobas 8000 ISE module part Numbers : 05964075001 cobas c 701 module part Numbers : 05641489001 cobas c 702 module part Numbers : 06473245001
- Status
- Terminated
- Initiated
- 2019-07-18
- Posted
- not on file
- Terminated
- 2020-06-15
- Reason
- Quality issue with high pressure solenoid valves
- Root cause
- Device Design
- Action
- 1. For the cobas 8000 ISE, cobas c 701, and cobas c 702 modules and the cobas c 513 analyzer, stop using these units immediately. Your FES will be in contact with you to replace these affected high pressure solenoid valves immediately. 2. In the event that the high pressure solenoid valves cannot be replaced immediately, for the cobas e 601 module, follow the wash station checks outlined in this Urgent Medical Device Correction (UMDC). Perform these checks once per shift. If these checks show favorable outcomes and quality control results are within range, patient results may be reported. If these checks produce unfavorable outcomes, stop using your cobas e 601 module and call the Roche Support Network Customer Support Center at 1-800-428-2336. 3. Complete the attached fax form and fax or email it according to the instructions on the form. 4. File this Urgent Medical Device Correction (UMDC) for future reference Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC.
- Quantity
- 19 units
- Distribution
- State NY MT HI IN MD TX OH KS KY CA FL OR LA MO AK
- Lot, serial or model codes
- cobas c 701 module Serial # : 19P2-01, 1390-03, 1139-04, 1139-05, 1139-06. cobas c 702 module serial # : 1126-03, 1237-10, 14C2-04 cobas¿ 8000 ISE module serial # : 1943-01, 1943-02, 1943-03, 1943-04, 1943-05, 1943-06, 1943-07, 1943-08, 1943-09, 1943-10, 15B3-05.
- 510(k) numbers
- ["K100853"]
- PMA numbers
- not on file
- Product code
- JJE
- Firm FEI
- 1823260
- Firm city
- Indianapolis
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28