FDA Medical Device Recalls (openFDA)
Nidek Inc: OPD-Scan III Refractive Power/Corneal Analyzer Opthalmic Software versions 1.00.08 1.05.07 1.10.01, 1.01.02 1.06.02 1.11.02, 1.02.01 1.07.01 1.12.03, 1.03.02 1.08.01 1.13.01, 1.04.03 1.09.01. Opthalmic: The OPD-Scan III is a diagnostic instrument that is indicated for use for Mapping of refractive error distribution of the eye by measurement and analysis of spherical power, cylindrical power, and cylinder axis.
- Recall number
- Z-2477-2015
- Recalling firm
- Nidek Inc
- Product
- OPD-Scan III Refractive Power/Corneal Analyzer Opthalmic Software versions 1.00.08 1.05.07 1.10.01, 1.01.02 1.06.02 1.11.02, 1.02.01 1.07.01 1.12.03, 1.03.02 1.08.01 1.13.01, 1.04.03 1.09.01. Opthalmic: The OPD-Scan III is a diagnostic instrument that is indicated for use for Mapping of refractive error distribution of the eye by measurement and analysis of spherical power, cylindrical power, and cylinder axis.
- Status
- Terminated
- Initiated
- 2015-06-15
- Posted
- 2015-08-27
- Terminated
- 2016-04-25
- Reason
- Software bug was found where there was no difference in Total and Corneal high-order aberrations, but differences were found in Internal high-order aberrations.
- Root cause
- Software design
- Action
- Dear Doctor letters mailing started on July 15, 2015 - first 100. They will be mailed out as the correction is received from Nidek, Japan. Dear Doctor letter will include software upgrade kit including CD or USB, instruction for upgrade and pre-paid response card to be returned to Nidek Inc.
- Quantity
- 3836
- Distribution
- Nationwide Distribution.
- Lot, serial or model codes
- Software versions: 1.00.08 1.05.07 1.10.01, 1.01.02 1.06.02 1.11.02, 1.02.01 1.07.01 1.12.03, 1.03.02 1.08.01 1.13.01, 1.04.03 1.09.01.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- HKO
- Firm FEI
- 2936921
- Firm city
- Fremont
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28