FDA Medical Device Recalls (openFDA)
Diagnostica Stago, Inc.: STA - Fibrinogen 5. Model/Catalog Number: 00674. The STA - Fibrinogen 5 kit is intended for use with the analyzers of the STA-R, STA Compact and STA Satellite families for the quantitative determination of fibrinogen levels in human citrated plasma by the clotting method of Clauss.
- Recall number
- Z-2476-2025
- Recalling firm
- Diagnostica Stago, Inc.
- Product
- STA - Fibrinogen 5. Model/Catalog Number: 00674. The STA - Fibrinogen 5 kit is intended for use with the analyzers of the STA-R, STA Compact and STA Satellite families for the quantitative determination of fibrinogen levels in human citrated plasma by the clotting method of Clauss.
- Status
- Open, Classified
- Initiated
- 2025-07-21
- Posted
- 2025-08-28
- Terminated
- not on file
- Reason
- Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
- Root cause
- Mixed-up of materials/components
- Action
- Diagnostica Stago notified consignees on about 07/21/2025 via emailed letter titled URGENT: MEDICAL DEVICE RECALL (CORRECTION). Consignees were instructed to temporarily systematically run a quality control just before a batch of UFH or LMWH, if you run STA-LIQUID ANTI-Xa in UFH and LMWH applications on your STA SATELLITE analyzer. This request is being asked while Stago works to develop a permanent solution for the analyzer. Consignees were also requested to complete and return the Acknowledgement Form and to notify customers if affected units were transferred or further distributed.
- Quantity
- 829,800 units (245,200 US)
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, AZ, CA, CO, FL, GA, LA, MD, ME, MN, MO, MT, NC, OH, OK, OR, PA, PR, SD, VT, WI, WY and the countries of Algeria, Argentina, Austria, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, France, French Polynesie, Guadeloupe, India, Iran, Israel, Jordan, Korea, Kuwait, Lebanon, Malaysia, Mexico, Morocco, Netherlands, Oman, Paraguay, Peru, Republique Tcheque, Romania, Saudi Arabia, Serbia, South Africa, Switzerland, Taiwan, Thailand, Tunisia, Turquie, United Arab Emirates, Uruguay, Vietnam.
- Lot, serial or model codes
- Model/Catalog Number: 00674. UDI-DI: 13607450006749
- 510(k) numbers
- ["K840211"]
- PMA numbers
- not on file
- Product code
- KQJ
- Firm FEI
- 2245451
- Firm city
- Parsippany
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28