FDA Medical Device Recalls (openFDA)
Customed, Inc: Laparotomy Pack, 900-2516, pack contains: ( I) TABLE COVER44" X 90" ( I) CAUTERY PENCIL PUSH BOTTOM (4) TOWELS ABSORBENT 15" X 20" L/F ( I) BULB SYRINGE 60cc LIF (4) DRAPE UTILITY WIT APE LIF (I) SKIN MARKER W/RULER ( I) SUTURE BAG FLORAL LIF ( I) Y ANKAUER SUCTION TUBE W/0 VENT LIF (2) COUNTER NDL /BLADE IOc MAG/CLEAR ( I) GOWN XL SMS IMPERV. REINFORCED LIF (5) LAP SPONGE PREWASH 18" X 18" XRD LIF ( I) GOWN SURG REINFORCED. LGE T/WRAP LIF ( I) TUBE SUCTION CONNECT W' X 12' LIF (2) LITE GLOVE LIF ( 10) GAUZE SPONGE4" X 4" 16PLY XRD LIF (2) DRAPE SHEET 41" X 58" SMS LIF ( I) CAUTERY TIP POLISHER LIF ( I) MERLING SKIN PREP APPLIC. 39ML LIF ( I) UTILITY BOWL PINT 16oz LIF ( I) SKIN STAPLE 35 WIDE LIF ( I) UTILITY BOWL QUART 32oz LIF ( I) BLADE SURGICAL II CARBON STEEL ( I) TRAY MAYO SMALL ( I) BLADE SURGICAL 20 CARBON STEEL ( I) MAYO STAND COVER REINFORCED LIF ( I) DRAPE LAPAROTOMY 100" X 72" X 124" W/POUCH Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
- Recall number
- Z-2476-2014
- Recalling firm
- Customed, Inc
- Product
- Laparotomy Pack, 900-2516, pack contains: ( I) TABLE COVER44" X 90" ( I) CAUTERY PENCIL PUSH BOTTOM (4) TOWELS ABSORBENT 15" X 20" L/F ( I) BULB SYRINGE 60cc LIF (4) DRAPE UTILITY WIT APE LIF (I) SKIN MARKER W/RULER ( I) SUTURE BAG FLORAL LIF ( I) Y ANKAUER SUCTION TUBE W/0 VENT LIF (2) COUNTER NDL /BLADE IOc MAG/CLEAR ( I) GOWN XL SMS IMPERV. REINFORCED LIF (5) LAP SPONGE PREWASH 18" X 18" XRD LIF ( I) GOWN SURG REINFORCED. LGE T/WRAP LIF ( I) TUBE SUCTION CONNECT W' X 12' LIF (2) LITE GLOVE LIF ( 10) GAUZE SPONGE4" X 4" 16PLY XRD LIF (2) DRAPE SHEET 41" X 58" SMS LIF ( I) CAUTERY TIP POLISHER LIF ( I) MERLING SKIN PREP APPLIC. 39ML LIF ( I) UTILITY BOWL PINT 16oz LIF ( I) SKIN STAPLE 35 WIDE LIF ( I) UTILITY BOWL QUART 32oz LIF ( I) BLADE SURGICAL II CARBON STEEL ( I) TRAY MAYO SMALL ( I) BLADE SURGICAL 20 CARBON STEEL ( I) MAYO STAND COVER REINFORCED LIF ( I) DRAPE LAPAROTOMY 100" X 72" X 124" W/POUCH Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
- Status
- Terminated
- Initiated
- 2014-05-20
- Posted
- 2014-08-29
- Terminated
- 2017-08-14
- Reason
- Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
- Root cause
- Package design/selection
- Action
- Customed Inc. sent a Recall Notification letter dated June 3, 2014 via email and followed by letter to all affected customers. The letter identified the affected products, problem and actions to be taken. The letter informed customers if the affected products are in their inventory to: 1. Remove products from sale (quarantine them) and cease distribution. 2. Complete and return the enclosed acknowledgement form to Customed Inc .Recall Coordinator. Affected products should be returned to Customed Inc for credit. Customers with questions were instructed to contact Customed Inc. Recall Coordinator at 1-787-801-0100 Ext. 7540
- Quantity
- 2 lots, 93 units, multiple packed units per lot
- Distribution
- Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
- Lot, serial or model codes
- Product code 900-2516, 2 lots: 131110250 131210792
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LRO
- Firm FEI
- 2648727
- Firm city
- Fajardo
- Firm state
- not on file
- Firm country
- PR
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28