FDA Medical Device Recalls (openFDA)
Zimmer, Inc.: 00789502601 VerSys Femoral Head Provisional 26mm -3.5 00789502602 VerSys Femoral Head Provisional 26mm +0 00789502603 VerSys Femoral Head Provisional 26mm +3.5 00789502604 VerSys Femoral Head Provisional 26mm +7 00789502605 VerSys Femoral Head Provisional 26mm +10.5 12/14 Neck Taper The provisional heads replicate final femoral implant geometry y and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.
- Recall number
- Z-2474-2012
- Recalling firm
- Zimmer, Inc.
- Product
- 00789502601 VerSys Femoral Head Provisional 26mm -3.5 00789502602 VerSys Femoral Head Provisional 26mm +0 00789502603 VerSys Femoral Head Provisional 26mm +3.5 00789502604 VerSys Femoral Head Provisional 26mm +7 00789502605 VerSys Femoral Head Provisional 26mm +10.5 12/14 Neck Taper The provisional heads replicate final femoral implant geometry y and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.
- Status
- Terminated
- Initiated
- 2011-04-01
- Posted
- 2012-09-28
- Terminated
- 2014-05-16
- Reason
- Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp
- Root cause
- Device Design
- Action
- Zimmer, Inc. sent an E-mail letter dated April 21, 2011, to all affected customers in the US. A June 2011 E-mail was sent to all customers outside the US. The letters requested Inventory & Operations Managers to check product for proper fit /function, complete the return form with worn provisional heads to Zimmer Inc, Attention Product Service, 1777 West Center St. Warsaw, IN 46580. For questions regarding this recall call 574-372-4753.
- Quantity
- 231, 641 all lots
- Distribution
- Worldwide Distribution - USA (nationwide) and internationally to Canada, Mexico, South America, Europe, Puerto Rico, China, and Korea.
- Lot, serial or model codes
- All lots distributed from lot 69998600 (2/20/1996) to present
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- HWT
- Firm FEI
- 1000220733
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28