Skip to the content

FDA Medical Device Recalls (openFDA)

Alcon Research, LTD.: This voluntary medical device correction is related to Alcon products that present a 23G nonvalved trocar cannula and infusion cannula together. These items are available as components within Accurus 23G TotalPLUS Pak, Accurus 23G Combined Procedure Packs (0.9mm and 1.1mm), Constellation 23G Total PLUS Pak, Constellation 23G Combined Pack (0.9 mm and 1.1 mm) and 23G Small Parts Kit. Single use 23G non-valved trocar cannulas are used to provide internal access to the ocular cavity during ophthalmic surgery. Single use infusion cannulas are used for the infusion of fluid or air during ophthalmic surgery to maintain intraocular pressure during posterior segment surgeries.

Recall number
Z-2470-2012
Recalling firm
Alcon Research, LTD.
Product
This voluntary medical device correction is related to Alcon products that present a 23G nonvalved trocar cannula and infusion cannula together. These items are available as components within Accurus 23G TotalPLUS Pak, Accurus 23G Combined Procedure Packs (0.9mm and 1.1mm), Constellation 23G Total PLUS Pak, Constellation 23G Combined Pack (0.9 mm and 1.1 mm) and 23G Small Parts Kit. Single use 23G non-valved trocar cannulas are used to provide internal access to the ocular cavity during ophthalmic surgery. Single use infusion cannulas are used for the infusion of fluid or air during ophthalmic surgery to maintain intraocular pressure during posterior segment surgeries.
Status
Terminated
Initiated
2012-05-29
Posted
2012-09-27
Terminated
2013-02-12
Reason
This medical device correction is related to Alcon products that present a 23G non-valved trocar cannula and infusion cannula together because of an issue regarding the loose fit between the 23G non-valved trocar cannula and infusion cannula that could potentially cause a detachment between the non-valved trocar cannula and infusion cannula during surgery.
Root cause
Device Design
Action
Alcon sent an Urgent a Voluntary Medical Device Correction letter on May 9, 2012. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed NOT to use the affected product, instead customers should use a 1-Count Valved Trocar Cannula and to complete and immediately return the Medical Device Correction Response Form to Alcon via fax at: (817) 916-9087 or email to: ReplacementCannulas@alconlabs.com. For questions customers should call 1-800-862-5266.
Quantity
93,319 (30, 413 distributed domestically & 62,906 internationally)
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Saudi Arabia, United Kingdom, Greece, Singapore, Germany, Switzerland, Kuwait, United Arabian Emirates, and Austria.
Lot, serial or model codes
1) 23G TotalPLUS Pak: Accurus Standalone - 8065750828, 2) 23G Combined Pack 0.9mm: Accurus Standalone - 8065750814, 3) 23G Combined Pack 1.1mm: Accurus Standalone - 8065750816, 4)23G TotalPLUS Pak: Constellation Standalone - 8065751058, 5) 23G Combined Pack 0.9mm: Constellation Standalone -8065751077, 6) 23G Combined Pack 1.1mm: Constellation Standalone - 8065751065. 7) 23G Small Parts Kit - 8065750842. Complete list of lot numbers will be provided with recall recommendation.
510(k) numbers
not on file
PMA numbers
not on file
Product code
NGY
Firm FEI
1000117501
Firm city
Houston
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register