Linvatec Corp. dba ConMed Linvatec: Catheter Irrigation Set labeled in part: ***24K Arthroscopy Outflow/Suction Tubing Set*** This Arthroscopy Tubing Set is intended for use, in conjunction with the ConMed Linvatec 24k Pump, for pump-induced outflow and/or suction of fluid from the joint during arthroscopic procedures.

Recall number
Z-2469-2012
Recalling firm
Linvatec Corp. dba ConMed Linvatec
Product
Catheter Irrigation Set labeled in part: ***24K Arthroscopy Outflow/Suction Tubing Set*** This Arthroscopy Tubing Set is intended for use, in conjunction with the ConMed Linvatec 24k Pump, for pump-induced outflow and/or suction of fluid from the joint during arthroscopic procedures.
Status
Terminated
Initiated
2011-07-01
Posted
2012-09-27
Terminated
2013-07-25
Reason
ConMed Linvatec recalled their 24k Arthroscopy Outflow/Suction Tubing Set (Product Number 24k100)because there is a possibility that the product may have a breach in the Tyvek seal that could potentially compromise the sterility of the contents.
Root cause
Employee error
Action
ConMed Linvatec sent a Urgent Medical Device Recall Notification via FedEx to all affected customers. This notice should be received July 5, 2011. The letter identified the product the problem and the action needed to be taken by the customer. All returned product under this removal action will be identified, segregated, and quarantined per ConMed Utica, N.Y. procedures. Returned product will be dispositioned in accordance with ConMed Utica, N.Y. procedures. This can include repackaged and resterilization with new lot number or product scrap. ConMed Utica, N.Y. will maintain control of all individual lot numbers returned and dispositioned. Firm is striving for 1005 effectiveness. If a Reply form is not received within approximately 30 days of the first mailing, ConMed Linvatec will make two additional contacts by mail approximately 30-35 days apart. Should you have any additional questions please contact our Customer Service at 800-237-0169 or fax at 727-319-5701.
Quantity
30064 units
Distribution
Worldwide Distribution--USA (nationwide) including the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NV, NY, OH, OR, PA, SC, TN, TX, UT, WA ,WI and Hawaii., and the countries of Australia Canada, China, Switzerland, Germany, Czech Republic, Denmark, Egypt, Spain, France, South Africa, United Kingdom, Israel, Italy, Netherlands, New Zealand, Poland, Portugal, Russia and Saudi Arabia.
Lot, serial or model codes
Catalog # 24k100 Lot #: 20853-000 20854-000 20855-000 21791-000 22172-000 22414-000 22513-000 22667-000 22724-000 22725-000 23606-000 23912 23912-000 23971-000 24177-000 24296-000 24348-000 24373-000 24476-000 24559-000 24671-000 24703-000 24905-000 25668-000 25794-000 25931-000 26025-000 26041-000 26294-000 26448-000 26486-000 27130-000 27255 27255-000 27322 27520
510(k) numbers
not on file
PMA numbers
not on file
Product code
GBX
Firm FEI
1017294
Firm city
Largo
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register