Skip to the content

FDA Medical Device Recalls (openFDA)

LeMaitre Vascular, Inc.: The Chevalier Valvulotome is a device used to disrupt the valves of veins which are to be used as grafts in the in situ or non-reversed bypass procedure. The Chevalier Valvulotome consists of a 70 cm stainless steel shaft with two tips: 1) the cutting tip is shaped like an inverted tulip with four teeth at its base. This tip is designed to cut the vein valve cusps. 2) the other tip is atraumatic and used to test the patency and the course of the vein before the valvulotomy. The Chevalier Valvulotome is available in three tip diameters: 2.5 mm (Catalog # 40025); 3.0 mm (Catalog # 40030); and 3.5 mm (Catalog # 40035)

Recall number
Z-2468-2023
Recalling firm
LeMaitre Vascular, Inc.
Product
The Chevalier Valvulotome is a device used to disrupt the valves of veins which are to be used as grafts in the in situ or non-reversed bypass procedure. The Chevalier Valvulotome consists of a 70 cm stainless steel shaft with two tips: 1) the cutting tip is shaped like an inverted tulip with four teeth at its base. This tip is designed to cut the vein valve cusps. 2) the other tip is atraumatic and used to test the patency and the course of the vein before the valvulotomy. The Chevalier Valvulotome is available in three tip diameters: 2.5 mm (Catalog # 40025); 3.0 mm (Catalog # 40030); and 3.5 mm (Catalog # 40035)
Status
Open, Classified
Initiated
2023-07-13
Posted
2023-08-28
Terminated
not on file
Reason
Burrs (molding flash) on the cutter may tear the harvested vien
Root cause
Nonconforming Material/Component
Action
LeMaitre issued Field Safety Notice (FSN) Chevalier Valvulotome (FSCA Ref: CAPA 2023-022) on July 13, 2023 to OUS Customers. Letter states reason for recall, health risk and action to take: Please quarantine the affected devices. Then, return the faxback/reply form (at the end of this FSN) to LeMaitre Vascular as soon as possible. Then, LeMaitre will give you instructions on how to return the devices.
Quantity
225 units
Distribution
International distribution to the countries of Belgium, France, Germany, and United Kingdom.
Lot, serial or model codes
(1)Catalog # 40025 UDI-DI: 00840663110469; (2)Catalog # 40030 UDI-DI: 00840663110476; (3) Catalog # 40035 UDI-DI: 00840663110483 Lot Numbers: All lots starting with CHV. Only lots beginning with CHV are being recalled. Older lots (with CL lot prefixes) are not being recalled. REF (Catalog #) LOT Catalog Number: (1)40025 Lot Number: CHV1003, CHV1006, CHV1022 (2) 40030 Lot Number: CHV1002, CHV1004, (3) 40035 Lot Number: CHV1001, CHV1005
510(k) numbers
not on file
PMA numbers
not on file
Product code
MGZ
Firm FEI
1220948
Firm city
Burlington
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register