Skip to the content

FDA Medical Device Recalls (openFDA)

Sunrise Medical (US) LLC: Quickie IRIS Wheelchair. model EIZ4-2 in combination with option code 183M02 - MONO Backrest system with Dynamic Backrest option. Provide mobility to persons limited to a sitting position.

Recall number
Z-2467-2015
Recalling firm
Sunrise Medical (US) LLC
Product
Quickie IRIS Wheelchair. model EIZ4-2 in combination with option code 183M02 - MONO Backrest system with Dynamic Backrest option. Provide mobility to persons limited to a sitting position.
Status
Terminated
Initiated
2015-07-27
Posted
2015-08-26
Terminated
2015-12-10
Reason
Quickie IRIS and Zippie IRIS tilt-in-space wheelchairs with MONO Backrest System with Dynamic Backrest may break over time resulting in a fall or injury to occupant.
Root cause
Device Design
Action
Sunrise Medical sent an Urgent Medical Device Field Correction letter dated July 27, 2015, to all affected dealers. The letter identified the product, the problem, and the action to be taken by the dealer. Dealers were instructed to immediately contact their customers to make arrangements to have the correction made with a replacement kit which will be supplied by Sunrise Medical to each dealer. Each dealer is requested to send back to Sunrise Medical the Acknowledgment and Response Form(s) by fax, email or regular mail once the work is completed. Customers with questions were instructed to contact Sunrise Medical Regulatory Affairs at (888) 208-4901.
Quantity
64 total wheelchairs, both models.
Distribution
Worldwide Distribution - US (nationwide) and Internationally to Australia, Germany and Canada.
Lot, serial or model codes
Model EIZ4-2, serial number range - IRS-046185-IRS-063437 and IRSE-058994 to IRSE-063928.
510(k) numbers
["K123975"]
PMA numbers
not on file
Product code
IOR
Firm FEI
2937137
Firm city
Fresno
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register