FDA Medical Device Recalls (openFDA)

Hitachi Medical Systems America Inc: Hitachi HIVISION Preirus/Avius Ultrasound equipped with V03.02 Software - Doppler Trace Measurement program. 510 (k) Numbers: Preirus - K093466 (6/1/10), Avius - K102901 (10/27/10). The Preirus and Avius are intended for use by trained personnel (doctors, sonographers, etc) for the diagnostic ultrasound evaluation of Abdominal, Cardiac, Intra-operative, Feta, Pediatric, Small Organ, Peripheral vessel, Biopsy, Trans-rectal, Trans-vaginal, Musculoskeletal, Neonatal Cephalic, Adult Cephalic, Endoscopy, Intra-luminal, Genecology, Urology and Laparoscopic clinic applications. The Modes of Operation of the HIVISION Preirus are B mode, M mode PW mode (Pulsed Wave Doppler), CW mode (Continuous Wave Doppler) Color Doppler, Amplitude Doppler (Color Flow Angiography), TDI (Tissue Doppler Imaging), 3D Imaging, 4D Imaging, RealTime Tissue Elastography, and Real Time Virtual Sonography.

Recall number
Z-2467-2011
Recalling firm
Hitachi Medical Systems America Inc
Product
Hitachi HIVISION Preirus/Avius Ultrasound equipped with V03.02 Software - Doppler Trace Measurement program. 510 (k) Numbers: Preirus - K093466 (6/1/10), Avius - K102901 (10/27/10). The Preirus and Avius are intended for use by trained personnel (doctors, sonographers, etc) for the diagnostic ultrasound evaluation of Abdominal, Cardiac, Intra-operative, Feta, Pediatric, Small Organ, Peripheral vessel, Biopsy, Trans-rectal, Trans-vaginal, Musculoskeletal, Neonatal Cephalic, Adult Cephalic, Endoscopy, Intra-luminal, Genecology, Urology and Laparoscopic clinic applications. The Modes of Operation of the HIVISION Preirus are B mode, M mode PW mode (Pulsed Wave Doppler), CW mode (Continuous Wave Doppler) Color Doppler, Amplitude Doppler (Color Flow Angiography), TDI (Tissue Doppler Imaging), 3D Imaging, 4D Imaging, RealTime Tissue Elastography, and Real Time Virtual Sonography.
Status
Terminated
Initiated
2011-04-18
Posted
2011-06-08
Terminated
2013-02-19
Reason
Software anomaly. The Doppler Trace Measurement feature on certain models of the firm's diagnostic ultrasound systems, may, under certain conditions whereby multiple measurements are done sequentially, record incorrect values. The recorded value is displayed to the user on a report; however depending upon the measurement area, displayed velocity values could be lower than at the true velocity lead
Root cause
Software design
Action
The firm, Hitachi, sent a "DEVICE CORRECTION NOTICE" letter dated April 15, 2011 to its customers. The letter describes the product, problem and actions to be taken. The customers were inform of the nature of the software error in the Doppler Trace Measurement feature; the circumstances under which this error can occur; and the safety issue involved should incorrect values generated lead to a false diagnosis of stenosis in a patient. The letter provides the customer with a workaround solution in order to avoid this issue. The customer were also instructed to call the HMSA Applications Helpline at 877-871-2777 to reprogram their system to temporarily eliminate the noted safted issue and to immediately have all selections unchecked on the Doppler trace measurement result store setting set up screen. The notification letter also states that Hitachi Medical Systems America (HMSA) will be having their service personnel schedule a visit at each customer location and install a software upgrade on each affected system unit in order to directly address, and permanently resolve this issue. If you need any further information regarding this notice, please contact the Manager of Regulatory Affairs at 800-800-3106 x3720 or email to: thistlewaited@hitachimed.com.
Quantity
24 system units
Distribution
Nationwide distribution: USA including states of: CA, MD, KY, FL, MN, NJ, IA, OH, TX, IL, and IA.
Lot, serial or model codes
All codes manufactured from 11/2/2009 to 3/4/2011, are subject to recall. Preirus (KE10584001-KE17585003)/Avius (KE11088001-KE11476102)
510(k) numbers
["K093466","K102901"]
PMA numbers
not on file
Product code
IYN
Firm FEI
1528028
Firm city
Twinsburg
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register