FDA Medical Device Recalls (openFDA)
Hitachi Medical Systems America Inc: Hitachi HIVISION Preirus/Avius Ultrasound equipped with V03.02 Software - Doppler Trace Measurement program. 510 (k) Numbers: Preirus - K093466 (6/1/10), Avius - K102901 (10/27/10). The Preirus and Avius are intended for use by trained personnel (doctors, sonographers, etc) for the diagnostic ultrasound evaluation of Abdominal, Cardiac, Intra-operative, Feta, Pediatric, Small Organ, Peripheral vessel, Biopsy, Trans-rectal, Trans-vaginal, Musculoskeletal, Neonatal Cephalic, Adult Cephalic, Endoscopy, Intra-luminal, Genecology, Urology and Laparoscopic clinic applications. The Modes of Operation of the HIVISION Preirus are B mode, M mode PW mode (Pulsed Wave Doppler), CW mode (Continuous Wave Doppler) Color Doppler, Amplitude Doppler (Color Flow Angiography), TDI (Tissue Doppler Imaging), 3D Imaging, 4D Imaging, RealTime Tissue Elastography, and Real Time Virtual Sonography.
- Recall number
- Z-2467-2011
- Recalling firm
- Hitachi Medical Systems America Inc
- Product
- Hitachi HIVISION Preirus/Avius Ultrasound equipped with V03.02 Software - Doppler Trace Measurement program. 510 (k) Numbers: Preirus - K093466 (6/1/10), Avius - K102901 (10/27/10). The Preirus and Avius are intended for use by trained personnel (doctors, sonographers, etc) for the diagnostic ultrasound evaluation of Abdominal, Cardiac, Intra-operative, Feta, Pediatric, Small Organ, Peripheral vessel, Biopsy, Trans-rectal, Trans-vaginal, Musculoskeletal, Neonatal Cephalic, Adult Cephalic, Endoscopy, Intra-luminal, Genecology, Urology and Laparoscopic clinic applications. The Modes of Operation of the HIVISION Preirus are B mode, M mode PW mode (Pulsed Wave Doppler), CW mode (Continuous Wave Doppler) Color Doppler, Amplitude Doppler (Color Flow Angiography), TDI (Tissue Doppler Imaging), 3D Imaging, 4D Imaging, RealTime Tissue Elastography, and Real Time Virtual Sonography.
- Status
- Terminated
- Initiated
- 2011-04-18
- Posted
- 2011-06-08
- Terminated
- 2013-02-19
- Reason
- Software anomaly. The Doppler Trace Measurement feature on certain models of the firm's diagnostic ultrasound systems, may, under certain conditions whereby multiple measurements are done sequentially, record incorrect values. The recorded value is displayed to the user on a report; however depending upon the measurement area, displayed velocity values could be lower than at the true velocity lead
- Root cause
- Software design
- Action
- The firm, Hitachi, sent a "DEVICE CORRECTION NOTICE" letter dated April 15, 2011 to its customers. The letter describes the product, problem and actions to be taken. The customers were inform of the nature of the software error in the Doppler Trace Measurement feature; the circumstances under which this error can occur; and the safety issue involved should incorrect values generated lead to a false diagnosis of stenosis in a patient. The letter provides the customer with a workaround solution in order to avoid this issue. The customer were also instructed to call the HMSA Applications Helpline at 877-871-2777 to reprogram their system to temporarily eliminate the noted safted issue and to immediately have all selections unchecked on the Doppler trace measurement result store setting set up screen. The notification letter also states that Hitachi Medical Systems America (HMSA) will be having their service personnel schedule a visit at each customer location and install a software upgrade on each affected system unit in order to directly address, and permanently resolve this issue. If you need any further information regarding this notice, please contact the Manager of Regulatory Affairs at 800-800-3106 x3720 or email to: thistlewaited@hitachimed.com.
- Quantity
- 24 system units
- Distribution
- Nationwide distribution: USA including states of: CA, MD, KY, FL, MN, NJ, IA, OH, TX, IL, and IA.
- Lot, serial or model codes
- All codes manufactured from 11/2/2009 to 3/4/2011, are subject to recall. Preirus (KE10584001-KE17585003)/Avius (KE11088001-KE11476102)
- 510(k) numbers
- ["K093466","K102901"]
- PMA numbers
- not on file
- Product code
- IYN
- Firm FEI
- 1528028
- Firm city
- Twinsburg
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28