KCI USA, Inc.: BariAir Therapy System The BariAir Therapy System is a risk management treatment system designed for large and / or difficult-to-manage patients. The BariAir Therapy System provides low air-loss pressure management therapy, pulsation, percussion, Turn Assist. flexible patient positioning capabilities and built-in scales for patients weighing 300 Ib* to 850 Ib ([136*-386 kg] including accessories). The BariAir Therapy System is indicated to aid in: " Patients whose body weight and size pose a significant risk or care management issue to the patient or staff during the performance of routine nursing care. " Large patients weighing between 300 Ib* and 850 Ib (136*-386 kg) including accessories. " Large patients who are difficult to turn. " Preventing and treating pressure ulcers. " Large patients requiring percussion therapy.

Recall number
Z-2466-2012
Recalling firm
KCI USA, Inc.
Product
BariAir Therapy System The BariAir Therapy System is a risk management treatment system designed for large and / or difficult-to-manage patients. The BariAir Therapy System provides low air-loss pressure management therapy, pulsation, percussion, Turn Assist. flexible patient positioning capabilities and built-in scales for patients weighing 300 Ib* to 850 Ib ([136*-386 kg] including accessories). The BariAir Therapy System is indicated to aid in: " Patients whose body weight and size pose a significant risk or care management issue to the patient or staff during the performance of routine nursing care. " Large patients weighing between 300 Ib* and 850 Ib (136*-386 kg) including accessories. " Large patients who are difficult to turn. " Preventing and treating pressure ulcers. " Large patients requiring percussion therapy.
Status
Terminated
Initiated
2012-09-17
Posted
2012-09-27
Terminated
2013-07-29
Reason
KCI has issued a medical device correction for the BariAir Therapy System Hand Pendent. The BariAir Therapy System is not designed to allow the patient to place the bed into the patient exit position. KCI has received complaints indicating that patients used the hand pendent to inadvertently place the BariAir Therapy System into a pseudo-exit position. In a small number of instances, this allow
Root cause
Device Design
Action
KCI sent an Urgent - Voluntary Medical Device Correction letter dated September 13, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers would be contacted by KCI to coordinate a shipment of a replacement product. For questions customers would instructed to call 1-800-275-4524, Select Option 3 followed by Option 2. For questions regarding this recall call 1-800-275-4524.
Quantity
575
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Saudi Arabia, United Kingdom, Singapore, Germany, Switzerland, Kuwait, United Arabian Emirates, and Austria.
Lot, serial or model codes
Multiple Serial Numbers with Model Numbers: 404000,405500, 405500RF
510(k) numbers
not on file
PMA numbers
not on file
Product code
FNL
Firm FEI
1625774
Firm city
San Antonio
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register