Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Healthcare Inc.: Philips Digital Diagnost-Mobile Detector Holder for Digital Diagnost (Single Detector, Dual Detector, Release 3) System codes: 712020, 712022, 712025 This system is used for making X-ray exposures for diagnostics.

Recall number
Z-2465-2012
Recalling firm
Philips Healthcare Inc.
Product
Philips Digital Diagnost-Mobile Detector Holder for Digital Diagnost (Single Detector, Dual Detector, Release 3) System codes: 712020, 712022, 712025 This system is used for making X-ray exposures for diagnostics.
Status
Terminated
Initiated
2012-09-06
Posted
2012-09-27
Terminated
2017-01-17
Reason
The Wireless Portable Detector (WPD) is not fixed securely inside the Mobile Detector Holder and may fall down
Root cause
Device Design
Action
Philips Healthcare sent a Urgent Field Safety Notice dated August 13, 2012, to all affected customers. Notifications began the week of September 6, 2012, for US and Worldwide Customers. The letter identified the product the problem and the action needed to be taken by the customer. Philips field service engineer will Secure the 4 countersunk screws, which are used to attach the lock mechanism, with thread locker (loctite). Further information or support concerning this issue, contact your local Philips representative. 1-800-722-9377.
Quantity
93 Units
Distribution
Worldwide Distribution--USA (nationwide) and the countries of Australia, Austria, Canada, Chile, Finland, France, Germany, Hungary. Italy, Japan, Latvia, Netherlands, New Zealand. Oman, South Korea, Switzerland and United Kingdom.
Lot, serial or model codes
Serial Numbers: SN10000053 SN11000644 SN10000648 SN10000649 SN11000226 SN10000703 SN11000032 SN10000407 SN11000105 SN11000112 SN11000212 SN11000231 SN11000030 SN10000468 SN11000146 SN11000217 SN11000238 SN11000407 SN11000154 SN10000297 SN10000515 SN11000129 SN11000266 SN11000412 SN11000442 SN11000063 SN11000441 SN10000416 SN10000414 SN10000424 SN11000034 SN10000018 SN11000050 SN10000113 SN10000442 SN11000057 SN11000419 SN10000482 SN10000670 SN10000690 SN10000689 SN10000694 SN11000461 SN10000544 SN11000095 SN11000329 SN11000464 SN10000554 SN10000635 SN10000107 SN10000413 SN10000557 SN11000092 SN11000100 SN11000118 SN11000145 SN11000161 SN11000193 SN11000194 SN11000228 SN11000294 SN11000371 SN10000038 SN10000101 SN10000102 SN10000641 SN10000664 SN11000027 SN10000537 SN10000650 SN11000130 SN11000142 SN11000216 SN11000269 SN11000286 SN11000288 SN11000311 SN11000320 SN10000363 SN10000246 SN10000512 SN11000033 SN11000053 SN11000080 SN11000271 SN11000478 SN11000474 SN11000013 SN11000005 SN11000008 SN11000004 SN11000141 SN11000007
510(k) numbers
["K982795"]
PMA numbers
not on file
Product code
MQB
Firm FEI
1218950
Firm city
Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register