Skip to the content

FDA Medical Device Recalls (openFDA)

American Contract Systems Inc: AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) HEART PACK - 205947, MODEL/ITEM NUMBER ANCV78BC; 2) VASCULAR PACK - 232024, MODEL/ITEM NUMBER ANVA82AB; 3) OFF PUMP CABG PACK - 170316, MODEL/ITEM NUMBER BHOP48O; 4) OFF PUMP CABG PACK - 170316, MODEL/ITEM NUMBER BHOP48P; 5) PEDIATRIC OPEN HEART PACK - 170212, MODEL/ITEM NUMBER BHPH18F; 6) PEDIATRIC OPEN HEART PACK - 170212, MODEL/ITEM NUMBER BHPH18G; 7) CABG ACCESSORY PACK, MODEL/ITEM NUMBER HGCB98F; 8) CV SET UP PACK, MODEL/ITEM NUMBER HGCV05K; 9) OPEN HEART PACK B - COMPONENTS, MODEL/ITEM NUMBER HGOH99H; 10) STRADLE TABLE PACK - 301659, MODEL/ITEM NUMBER MHCV10AM; 11) MINOR VASCULAR PACK, MODEL/ITEM NUMBER MTVS73D; 12) AV GRAFT RCH 0246952, MODEL/ITEM NUMBER RCAV40C; 13) OPEN HEART PART 1 RCH 0247159, MODEL/ITEM NUMBER RCOH61G; 14) CARDIO 612968, MODEL/ITEM NUMBER TNCA08T; 15) CV B PACK, MODEL/ITEM NUMBER TNCB10AM; 16) A-V FISTULA - 206042, MODEL/ITEM NUMBER UDAV16AD; 17) CAROTID PACK - 206045, MODEL/ITEM NUMBER UDCP24AF; 18) FEM POP PACK - 206052, MODEL/ITEM NUMBER UDFP36AH; 19) CARDIOVASCULAR SUPPLY (PS 907285), MODEL/ITEM NUMBER UICD62AN; 20) SFCH PEDS CARDIAC SUPPLY PACK (PS131653), MODEL/ITEM NUMBER UICS88S; 21) SFCH ECMO PACK (PS 107116), MODEL/ITEM NUMBER UIPD03Q; 22) THORACIC SETUP (PS 035700), MODEL/ITEM NUMBER UITC53AL; 23) VASCULAR PACK, MODEL/ITEM NUMBER UIVA75J; 24) CABG ACCESSORY PACK 175619, MODEL/ITEM NUMBER UMCA14E; 25) OPEN HEART ACCESS PACK 175628, MODEL/ITEM NUMBER UMOH42E; 26) OPEN HEART ACCESS PACK 175628, MODEL/ITEM NUMBER UMOH42F; 27) OPEN HEART ACCESS PACK 175628, MODEL/ITEM NUMBER UMOH42G.

Recall number
Z-2464-2025
Recalling firm
American Contract Systems Inc
Product
AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) HEART PACK - 205947, MODEL/ITEM NUMBER ANCV78BC; 2) VASCULAR PACK - 232024, MODEL/ITEM NUMBER ANVA82AB; 3) OFF PUMP CABG PACK - 170316, MODEL/ITEM NUMBER BHOP48O; 4) OFF PUMP CABG PACK - 170316, MODEL/ITEM NUMBER BHOP48P; 5) PEDIATRIC OPEN HEART PACK - 170212, MODEL/ITEM NUMBER BHPH18F; 6) PEDIATRIC OPEN HEART PACK - 170212, MODEL/ITEM NUMBER BHPH18G; 7) CABG ACCESSORY PACK, MODEL/ITEM NUMBER HGCB98F; 8) CV SET UP PACK, MODEL/ITEM NUMBER HGCV05K; 9) OPEN HEART PACK B - COMPONENTS, MODEL/ITEM NUMBER HGOH99H; 10) STRADLE TABLE PACK - 301659, MODEL/ITEM NUMBER MHCV10AM; 11) MINOR VASCULAR PACK, MODEL/ITEM NUMBER MTVS73D; 12) AV GRAFT RCH 0246952, MODEL/ITEM NUMBER RCAV40C; 13) OPEN HEART PART 1 RCH 0247159, MODEL/ITEM NUMBER RCOH61G; 14) CARDIO 612968, MODEL/ITEM NUMBER TNCA08T; 15) CV B PACK, MODEL/ITEM NUMBER TNCB10AM; 16) A-V FISTULA - 206042, MODEL/ITEM NUMBER UDAV16AD; 17) CAROTID PACK - 206045, MODEL/ITEM NUMBER UDCP24AF; 18) FEM POP PACK - 206052, MODEL/ITEM NUMBER UDFP36AH; 19) CARDIOVASCULAR SUPPLY (PS 907285), MODEL/ITEM NUMBER UICD62AN; 20) SFCH PEDS CARDIAC SUPPLY PACK (PS131653), MODEL/ITEM NUMBER UICS88S; 21) SFCH ECMO PACK (PS 107116), MODEL/ITEM NUMBER UIPD03Q; 22) THORACIC SETUP (PS 035700), MODEL/ITEM NUMBER UITC53AL; 23) VASCULAR PACK, MODEL/ITEM NUMBER UIVA75J; 24) CABG ACCESSORY PACK 175619, MODEL/ITEM NUMBER UMCA14E; 25) OPEN HEART ACCESS PACK 175628, MODEL/ITEM NUMBER UMOH42E; 26) OPEN HEART ACCESS PACK 175628, MODEL/ITEM NUMBER UMOH42F; 27) OPEN HEART ACCESS PACK 175628, MODEL/ITEM NUMBER UMOH42G.
Status
Open, Classified
Initiated
2025-07-14
Posted
2025-08-28
Terminated
not on file
Reason
During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.
Root cause
Under Investigation by firm
Action
Owens & Minor issued an "URGENT: MEDICAL DEVICE CORRECTION" notice to its consignees on 07/15/2025 via email. The notice explained the problem, potential risk to health, and requested the following: ACTIONS TO TAKE: " Discontinue use of all affected components. " Add warning labels to affected procedure trays to ensure the components are discarded. A labeling template is enclosed for printing labels which can be affixed on all affected kits in inventory. This label should be applied to each affected procedure tray and product case in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. o Customers (end-users): If you receive a product from a distributor, please contact the distributor or your sales representative with any questions you may have. o Distributors: If any of the affected procedure trays identified above were further distributed, please ensure all end-users are appropriately notified of this product correction and maintain records of effectiveness for your consignees. " Complete the attached Recall Response Form and return as soon as possible to GM-OMRA_RECALLS@owens-minor.com. Report any quality problems or adverse events associated with the products listed in this notification to complaints@owens-minor.com. If you have any questions, contact the Manager, Regulatory Affairs at 470-280-4277 or email to: GM-OMRA_RECALLS@owens-minor.com.
Quantity
31506 units
Distribution
US distribution to states of: IL and LA
Lot, serial or model codes
1) MODEL/ITEM NUMBER ANCV78BC: UDI/DI 00191072235169, Kit Lot Numbers: 42-7844511, 42-8133811, 42-8037211, 42-7952411, 42-7882011; 2) MODEL/ITEM NUMBER ANVA82AB: UDI/DI 00191072235183, Kit Lot Numbers: 42-8025011, 42-8108511, 42-7928511; 3) MODEL/ITEM NUMBER BHOP48O: UDI/DI 00191072221490, Kit Lot Numbers: 9207467, 8862838, 9140871, 9049946, 9007346, 8934280; 4) MODEL/ITEM NUMBER BHOP48P: UDI/DI 00191072234544, Kit Lot Numbers: 9268782, 9345847; 5) MODEL/ITEM NUMBER BHPH18F: UDI/DI 00191072224651, Kit Lot Numbers: 8993526, 9098058, 8899297; 6) MODEL/ITEM NUMBER BHPH18G: UDI/DI 00191072234551, Kit Lot Numbers: 9238544; 7) MODEL/ITEM NUMBER HGCB98F: UDI/DI 00191072205285, Kit Lot Numbers: 42-7813711; 8) MODEL/ITEM NUMBER HGCV05K: UDI/DI 00191072212474, Kit Lot Numbers: 42-8097611, 42-8021411, 42-7870011, 42-7831911, 42-8093211; 9) MODEL/ITEM NUMBER HGOH99H: UDI/DI 00191072235336, Kit Lot Numbers: 42-8119511, 42-8044311, 42-7874411; 10) MODEL/ITEM NUMBER MHCV10AM: UDI/DI 00191072235619. Kit Lot Numbers: 42-7848211; 11) MODEL/ITEM NUMBER MTVS73D: UDI/DI 00191072237583, Kit Lot Numbers: 12-7397911, 12-7335111, 12-7072111, 12-7142511, 12-7224911; 12) MODEL/ITEM NUMBER RCAV40C: UDI/DI 00191072235381, Kit Lot Numbers: 42-8053511, 42-8151011, 42-7978811, 42-7853911; 13) MODEL/ITEM NUMBER RCOH61G: UDI/DI 00191072229823, Kit Lot Numbers: 42-7803611, 42-7941711, 42-7997711, 42-8075711; 14) MODEL/ITEM NUMBER TNCA08T: UDI/DI 00191072235534, Kit Lot Numbers: 42-7827011, 42-8158811; 15) MODEL/ITEM NUMBER TNCB10AM: UDI/DI 00191072232991, Kit Lot Numbers: 42-7864811, 42-7971311, 42-8065311; 16) MODEL/ITEM NUMBER UDAV16AD: UDI/DI 00191072216229, Kit Lot Numbers: 42-7919811; 17) MODEL/ITEM NUMBER UDCP24AF: UDI/DI 00191072235305, Kit Lot numbers: 42-8010311, 42-8109811; 18) MODEL/ITEM NUMBER UDFP36AH: UDI/DI 00191072235329, Kit Lot Numbers: 42-7835011, 42-7902811, 42-8014011, 42-8091211; 19) MODEL/ITEM NUMBER UICD62AN: UDI/DI 00191072200495, Kit Lot Numbers: 42-7874211, 42-7934511, 42-7999011, 42-8072211, 42-8132211; 20) MODEL/ITEM NUMBER UICS88S: UDI/DI 00191072235428, Kit Lot Numbers: 42-7821511, 42-7913211, 42-7979011, 42-8032611, 42-8106011; 21) MODEL/ITEM NUMBER UIPD03Q: UDI/DI 00191072235435, Kit Lot Numbers: 42-8083011, 42-8154511; 22) MODEL/ITEM NUMBER UITC53AL: UDI/DI 00191072235572, Kit Lot Numbers: 42-7887011, 42-7993911, 42-8071411; 23) MODEL/ITEM NUMBER UIVA75J: UDI/DI 00191072235602, Kit Lot Numbers: 42-7994011, 42-8056011, 42-8114011, 42-7827411, 42-7916411; 24) MODEL/ITEM NUMBER UMCA14E: UDI/DI 00191072222510, Kit Lot Numbers: 334022, 9238536; 25) MODEL/ITEM NUMBER UMOH42E: UDI/DI 00191072222534, Kit Lot Numbers: 8860283; 26) MODEL/ITEM NUMBER UMOH42F: UDI/DI 00191072224712, Kit Lot Numbers: 9125072, 9044687, 8977422, 8925203, 9167965; 27) MODEL/ITEM NUMBER UMOH42G: UDI/DI 00191072234896, Kit Lot Numbers: 9202772, 9334223.
510(k) numbers
not on file
PMA numbers
not on file
Product code
OEZ
Firm FEI
3008492337
Firm city
Tiffin
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register