Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Siemens Magnetom Skyra, Verio, Spectra and Biograph MMR Nuclear Magnetic Resonance Imaging System and Tomographic Imager

Recall number
Z-2464-2012
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Siemens Magnetom Skyra, Verio, Spectra and Biograph MMR Nuclear Magnetic Resonance Imaging System and Tomographic Imager
Status
Terminated
Initiated
2012-09-12
Posted
2012-09-26
Terminated
2014-01-14
Reason
When positioning the Flex Large 4 Coil (part No. 8625761) off-center in left-right direction as it is used for example for hip and shoulder imaging and, in addition at an off-center position with respect to the head-foot direction, the coil can heat up in the area of the electronic housing (white plastic lids).
Root cause
Device Design
Action
Siemens Healthcare sent a Customer Safety Advisory Notice dated September 5, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. The letter included updated instruction to all customers. For questions regarding this recall call 610-219-6300.
Quantity
370
Distribution
Nationwide Distribution - USA including Washington, DC and AL, AK, AZ, AR, CA, CO, CT, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WI and Puerto Rico.
Lot, serial or model codes
Model numbers: 10276755 (Verio), 10684333 (Verio Dot), 10684334 (Verio Dot upgrade), 10432915 (Skyra), 10655588 (Spectra), and 10433372 (Biograph mMR)
510(k) numbers
["K072237","K101347","K103429","K121160"]
PMA numbers
not on file
Product code
LNH
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register