Skip to the content

FDA Medical Device Recalls (openFDA)

Imaging Sciences International, LLC: DEXIS Imaging Suite - catalog number 1.008.7932, version numbers 10.0.0 through 10.0.3 Radiological Image Processing System

Recall number
Z-2463-2012
Recalling firm
Imaging Sciences International, LLC
Product
DEXIS Imaging Suite - catalog number 1.008.7932, version numbers 10.0.0 through 10.0.3 Radiological Image Processing System
Status
Terminated
Initiated
2012-07-30
Posted
2012-09-26
Terminated
2016-10-13
Reason
A condition can occur during renumbering of the patient database using the DEXCopy feature. When renumbering is being performed using the DEXCopy feature and multiple patients with an identical name are present, the software will assign all subsequent records to the first patient record, resulting in mixed patient records. In order for this problem to occur several conditions need to be met: 1)
Root cause
Software design
Action
DEXIS sent a Medical Device Corrective Action letter dated July 30, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. An Acknowledgement Form was included with the notification that is to be returned by the consignee to the recalling firm. Customers were instructed to call DEXIS Technical Support at 888-883-3947 as soon as possible to have Technical Support review and potentially correct any issues that may have resulted from its use. To request a free-of-charge upgrade kit, customers should email a request to orders@dexis.com. For questions regarding this recall call 215-997-5666,
Quantity
366
Distribution
Worldwide Distribution - USA including Washington, DC, AL, AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MO, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI and the country of Canada.
Lot, serial or model codes
Catalog number 1.008.7932, version numbers 10.0.0 through 10.0.3
510(k) numbers
["K090461"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
3021551875
Firm city
Hatfield
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28