Skip to the content

FDA Medical Device Recalls (openFDA)

SCC Soft Computer: Product is SoftPath ASCII Software Releases 1.2, 2.1, 2.2, and 2.3.

Recall number
Z-2463-2008
Recalling firm
SCC Soft Computer
Product
Product is SoftPath ASCII Software Releases 1.2, 2.1, 2.2, and 2.3.
Status
Terminated
Initiated
2006-02-23
Posted
2008-09-22
Terminated
2008-11-20
Reason
In the creation of revised report and supplemental reports diagnosis, text was inserted from another case.
Root cause
Software design
Action
Consignees were notified of recall and provided with a software correction patch ICC number 3618.
Quantity
201
Distribution
Distributed to 106 consignees located throughout the United States and Canada.
Lot, serial or model codes
Version 1.2, 2.1, 2.2 and 2.3
510(k) numbers
not on file
PMA numbers
not on file
Product code
JQP
Firm FEI
1000307651
Firm city
Clearwater
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register