FDA Medical Device Recalls (openFDA)
Ferno-Washington Inc: Wheel caster set screw for Ferno iNX Wheeled Stretcher. Ferno iNX Wheeled Stretcher for pre-hospital patient transport.
- Recall number
- Z-2461-2015
- Recalling firm
- Ferno-Washington Inc
- Product
- Wheel caster set screw for Ferno iNX Wheeled Stretcher. Ferno iNX Wheeled Stretcher for pre-hospital patient transport.
- Status
- Terminated
- Initiated
- 2015-05-01
- Posted
- 2015-08-25
- Terminated
- 2015-11-06
- Reason
- The firm discovered an improperly installed set screw in one caster arm assembly on a unit in production. The screw assists in preventing the bearing cup on the caster assembly from backing out of the caster arm during use.
- Root cause
- Process control
- Action
- Ferno's first action of May 1, 2015 was to telephone all customers to alert them to the potential issue and explain our planned inspection. Ferno's Technical Support Manager emailed a copy of the Service Notice and iNX Equipment Evaluation Form to the customers as well.
- Quantity
- 26 units
- Distribution
- Distributed in the states of AZ, CO, IA, MD, OH, PA & WA, and the countries of Australia, Canada & United Kingdom.
- Lot, serial or model codes
- Serial numbers: 15F001703, 15F001704, 15F001705, 15F001706, 15F001707, 15F001708, 15F001709, 15F001710, 15F001711, 15F001712, 15F001713, 15F001714, 15F001715, 15F001716, 15F001717, 15F001719, 15F001720, 15F001721, 15F001722, 15F001723, 15F001724, 15F001725, 15F001726, 15F001727, 15F001729 & 15F001730.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FPO
- Firm FEI
- 1523574
- Firm city
- Wilmington
- Firm state
- OH
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as