Philips Medical Systems North America Co. Phillips: Nuclear Magnetic Resonance Imaging System (MRI). Trace item 9896-030-14351 -1.5T Sense XL Torso coil 12nc on coil: 4535-671-4188x.

Recall number
Z-2461-2010
Recalling firm
Philips Medical Systems North America Co. Phillips
Product
Nuclear Magnetic Resonance Imaging System (MRI). Trace item 9896-030-14351 -1.5T Sense XL Torso coil 12nc on coil: 4535-671-4188x.
Status
Terminated
Initiated
2008-09-30
Posted
2010-09-22
Terminated
2010-09-30
Reason
The cables of the coil may become too hot and cause burns when placed too close to the patient. While using the 1.5T Sense XL Torso coil, unwanted radiofrequencies (RF) interaction with the patient can occur. Especially in combination with high SAR levels and when the RF cables of the coil are too close to the patient, this RF interaction has resulted in RF burns to patients.
Root cause
Other
Action
Philips Healthcare issued URGENT-Device Correction Notices dated September 30, 2008 to their consignees. The notice identified the recalled device as 1.5 Intera and 1.5T Achieva MRI systems, described the risk to patients, and the actions to be taken by customers.. The firm has issued Field Change Order (FCO) 78100242. Field service representatives will visit each consignee and conduct a hardware upgrade, applying a sleeve over the cable that will ensure proper spacing between the cable and patient, reducing the chance of RF interaction and burns. The consignees should contact the Philips Call Center at 1-800-772-9377, #5 #3, #1 and reference "FCO 78100242 - PATIENT CABLE SLEEVE FOR 1.5T SENSE XL TORSO COIL".
Quantity
47 units total (Achieva 1.5T and Intera 1.5T)
Distribution
Nationwide Distribution: Throughout USA. No distribution to foreign consignees.
Lot, serial or model codes
Devices are identified as Site numbers: 544006, 537871, 534961, 544704, 545154, 545236, 545219, 545943, 546637, 545879, 545914, 543286, 547585, 548729, 550876, 553379, 550044, 554797, 41443789, 555412, 41444993, 41445232, 556803, 557348, 557344, 557513, 557179, 557644, 557560, 551656, 558845, 41446147, 552550, 552516, 555787, 41445636, 41455922, 83325, 558369, 86968, 101408, 101877, 102110, 102353, 102004, 104343, and 41455948.
510(k) numbers
["K052013"]
PMA numbers
not on file
Product code
MOS
Firm FEI
1000524572
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register